Z-0871-2010

PTS Panels Cholesterol test strips for self-testing or professional use with CardioChek analyzers...

Polymer Technology Systems / initiated 2009-04-02 / Terminated / root cause: manufacturing process

Polymer Technology Systems, Inc. began a recall of PTS Panels Cholesterol test strips for self-testing or professional use with CardioChek analyzers, Manufactured by Polymer Technology Systems, Inc... on . The reason FDA records is "The memory chip was coded with incorrect product information". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-0871-2010
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Polymer Technology Systems
Product
PTS Panels Cholesterol test strips for self-testing or professional use with CardioChek analyzers, Manufactured by Polymer Technology Systems, Inc., Indianapolis, IN.Catalog # 1711. Intended to provide a quantitative measurement of total cholesterol in whole blood.
Product code
CHH Enzymatic Esterase--Oxidase, Cholesterol
Reason for recall
The memory chip was coded with incorrect product information.
Root cause tag
manufacturing process
FDA root cause
Process control
Action
Polymer Technology Systems, Inc. issued an "Urgent - Field Correction" letter dated April 2, 2009. Consignees were advised of the problem and asked to identify how many kits were in inventory so that the firm could replace product. For further information, contact Polymer Technology Systems, Inc. Customer Service at 1-877-870-5610.
Quantity in commerce
486 vials
Distribution
Nationwide Distribution
Product codes and lots
Lot Number: C908.