Z-0871-2010
PTS Panels Cholesterol test strips for self-testing or professional use with CardioChek analyzers...
Polymer Technology Systems / initiated 2009-04-02 / Terminated / root cause: manufacturing process
Polymer Technology Systems, Inc. began a recall of PTS Panels Cholesterol test strips for self-testing or professional use with CardioChek analyzers, Manufactured by Polymer Technology Systems, Inc... on . The reason FDA records is "The memory chip was coded with incorrect product information". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.
Recall record
- Recall number
- Z-0871-2010
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Polymer Technology Systems
- Product
- PTS Panels Cholesterol test strips for self-testing or professional use with CardioChek analyzers, Manufactured by Polymer Technology Systems, Inc., Indianapolis, IN.Catalog # 1711. Intended to provide a quantitative measurement of total cholesterol in whole blood.
- Product code
- CHH Enzymatic Esterase--Oxidase, Cholesterol
- Reason for recall
- The memory chip was coded with incorrect product information.
- Root cause tag
- manufacturing process
- FDA root cause
- Process control
- Action
- Polymer Technology Systems, Inc. issued an "Urgent - Field Correction" letter dated April 2, 2009. Consignees were advised of the problem and asked to identify how many kits were in inventory so that the firm could replace product. For further information, contact Polymer Technology Systems, Inc. Customer Service at 1-877-870-5610.
- Quantity in commerce
- 486 vials
- Distribution
- Nationwide Distribution
- Product codes and lots
- Lot Number: C908.