Enpath Myopore Sutureless Myocardial Pacing Lead, Ref: 511211. the contents of the inner package have been...
Greatbatch Medical began a recall of Enpath Myopore Sutureless Myocardial Pacing Lead, Ref: 511211. the contents of the inner package have been sterilized by ethylene-oxide gas. Manufactured by... on . The reason FDA records is "The silicone tubing of affected leads was processed outside of validated parameters. Although the material remains biocompatible, test data regarding the performance characteristics are not available". As of , the recall status on record is Terminated.
Recall record
- Recall number
- Z-0581-2010
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Greatbatch Medical
- Product
- Enpath Myopore Sutureless Myocardial Pacing Lead, Ref: 511211. the contents of the inner package have been sterilized by ethylene-oxide gas. Manufactured by: Enpath Medical, Inc., 2300 Berkshire Lane North, Minneapolis, MN 55441 USA.
- Product code
- DTB Permanent Pacemaker Electrode
- Reason for recall
- The silicone tubing of affected leads was processed outside of validated parameters. Although the material remains biocompatible, test data regarding the performance characteristics are not available.
- FDA root cause
- Component change control
- Action
- Customers were contacted by phone beginning 09/01/09. They were asked to hold product and determine if they had distributed any product. A written notification followed which identified the product, described the issue and asked for product to be returned. Product was distributed to two consignees and contained the following day. Questions are directed to the company at 612-559-2613.
- Quantity in commerce
- 35 units
- Distribution
- CA and MN
- Product codes and lots
- Serial Numbers: 148399, 148400, 148401, 148402, 148403, 148404, 148405, 148406, 148407, 148408, 148409, 148410, 148411, 148412, 148413, 148414, 148415, 148416, 148417, 148418, 148419, 148420, 148421, 148422, 148423, 148424, 148427, 148430, 148431, 148433, 148434, 148435, 148436, 148437, 148438.