Z-1005-2011

10032746 Acuson Antares PE Product Version 5.0, manufactured by Siemens Medical Solutions, Mountain View, CA...

Siemens Medical Solutions USA, Inc. / initiated 2009-10-02 / Terminated / root cause: manufacturing process

Siemens Medical Solutions USA, Inc. began a recall of 10032746 Acuson Antares PE Product Version 5.0, manufactured by Siemens Medical Solutions, Mountain View, CA. Intended for Abdominal, Intraoperative, Small... on . The reason FDA records is "ECG waveform does not accurately align with 2D, M-Mode or Spectral Images". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-1005-2011
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Siemens Medical Solutions USA, Inc.
Product
10032746 Acuson Antares PE Product Version 5.0, manufactured by Siemens Medical Solutions, Mountain View, CA. Intended for Abdominal, Intraoperative, Small Parts, Transcranial, OB/Gyn, Cardiac, Transesophageal, Pelvic, Neonatal/Adult Cephalic, Vascular, Intravascular, Musculoskeletal, Superficial Musculoskeletal, Great Vessel, and Peripheral Vascular applications.
Product code
IYN System, Imaging, Pulsed Doppler, Ultrasonic
Reason for recall
ECG waveform does not accurately align with 2D, M-Mode or Spectral Images.
Root cause tag
manufacturing process
FDA root cause
Nonconforming Material/Component
Action
The firm's Customer Safety Advisory notice was mailed to all affected customers. The letter describes when the malfunction occurs and the potential risk. It also recommended that users collect corresponding evidence with spectral waveforms that would confirm or deny findings based on ECG, in order to avoid potential risks for misdiagnosis. The firm planned to exchange the faulty hardware component in the system. Customers would be contacted to schedule for the hardware exchange. Questions regarding this issue should be directed to the local service support person.
Quantity in commerce
67 units total
Distribution
Worldwide Distribution
Product codes and lots
Model number 10032746