Z-0605-2010

AMO Tecnis 1-Piece Intraocular Lens (Model ZCB00)

Advanced Medical Optics, Inc. / initiated 2009-11-11 / Terminated / root cause: manufacturing process

Abbott Medical Optics Inc (AMO) began a recall of AMO Tecnis 1-Piece Intraocular Lens (Model ZCB00) on . The reason FDA records is "Abbott Medical Optics (AMO) has received complaints of some Tecnis 1-Piece IOLs (model ZCB00) have adhered to the lens optic after insertion into the eye. This can lead to a need for additional manipulation to separate...". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-0605-2010
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Advanced Medical Optics, Inc.
Product
AMO Tecnis 1-Piece Intraocular Lens (Model ZCB00)
Product code
HQL Intraocular Lens
Reason for recall
Abbott Medical Optics (AMO) has received complaints of some Tecnis 1-Piece IOLs (model ZCB00) have adhered to the lens optic after insertion into the eye. This can lead to a need for additional manipulation to separate the haptics from the optic during surgery which could lead to an injury.
Root cause tag
manufacturing process
FDA root cause
Process design
Action
Beginning on Friday, November 6, 2009, AMO began distribution of the Recall Notification Letter via Federal Express to the 874 AMO Tecnis 1-piece IOL customer accounts in the US. Included with the Recall Notification letter is a facsimile delivery confirmation that each customer was instructed to fax back to AMO to confirm receipt of the Recall Notification Letter. --- Beginning on Monday, November 9, 2009, AMO began distribution of the Recall Notification letter to the AMO Tecnis 1-piece IOL customer/distributors accounts outside the USA. "THIS RECALL NOTIFICATION DOES NOT INVOLVE TECNIS 1-PIECE IOL THAT HAVE BEEN IMPLANTED. ONCE THE HAPTICS HAVE BEEN SUCCESSFULLY DEPLOYED, THESE LENSES FUNCTION PROPERLY." Questions should be directed to an AMO Customer Service Representative at 1-877-AMO-4LIFE (1-877-266-4543).
Quantity in commerce
246000 Lenses
Distribution
AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, ID, IL, IN, KS, KY, LA, MA, ME, MI, MN, MO, MS, NC, NH, NJ, NV, NY, OH, OK, OR, PA, RI, SC, TN, TX, UT, VA, VT, WA, WI, WV, WY, Hawaii, and Puerto Rico. INTERNATIONAL: Austria, Azerbaijan, Belgium, Canada, Switzerland, Cyprus, Czech Republic, Germany, Estonia, Spain, Finland, France, United Kingdom, Hungary, Ireland, Israel, Italy, Lebanon, Lichtenstein, Libya, Martinique, Netherlands, Poland, Palestine, Portugal, Saudi Arabic, Sweden, Slovenia, Turkey, and South Africa.
Product codes and lots
All Serial numbers of tecnis 1-Piece Intraocular Lenses that were processed through lens surface treatment at the AMO PR facility between May 19 and October 16, 2009 - with serial numbers ending in 0906, 0907, 0908, and 0909. and subset with serial numbers ending in 0905 and 0910.  Expiration/Use By Date (YYYY-MM): 2011-05, 2011-06, 2011-07, 2011-08, 2011-09, and 2011-10