Cybernet Medical MedStar Telemedicine System, Manufactured by Cybernet Medical, Ann Arbor, MI; Model 533-210...
Cybernet Systems Corp. began a recall of Cybernet Medical MedStar Telemedicine System, Manufactured by Cybernet Medical, Ann Arbor, MI; Model 533-210. The device is for a remote medical data... on . The reason FDA records is "A software problem renders these devices incapable of capturing lung function, oxygen saturation and pulse data collected from multiple tests in accordance with the instructions for use". As of , the recall status on record is Terminated, and MDM tags the root cause as design.
Recall record
- Recall number
- Z-0852-2010
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Cybernet Systems Co., Ltd.
- Product
- Cybernet Medical MedStar Telemedicine System, Manufactured by Cybernet Medical, Ann Arbor, MI; Model 533-210. The device is for a remote medical data collection device for the monitoring of patients by healthcare providers. It is intended to work with a weight scale, spirometer, blood pressure unit and pulse oximeter and to transmit data over a phone line.
- Product code
- DXH Transmitters And Receivers, Electrocardiograph, Telephone
- Reason for recall
- A software problem renders these devices incapable of capturing lung function, oxygen saturation and pulse data collected from multiple tests in accordance with the instructions for use.
- Root cause tag
- design
- FDA root cause
- Component design/selection
- Action
- The firm sent out a recall letter dated 10/27/09. Consignees were asked to return the devices in prepaid shipping containers. The recalling firm would do a software upgrade and return the device to the consignee. The firm expects it to take 3-6 months to upgrade all devices. Contact Cybernet Medically immediately, at 1-800-CYBERNET, with the serial number and code versions of the devices in question and Cybernet will start the RMA process for returning the devices for the updates, free of cost.
- Quantity in commerce
- 60
- Distribution
- Nationwide distribution: California, Florida, Georgia, Illinois, Louisiana, Michigan, Mississippi, New York, Oklahoma and West Virginia.
- Product codes and lots
- Serial number range affected was 2778 through 4085. Serial numbers 2826, 3719, 3726, 4045 and 4048 have not yet been corrected.