ERIC (Electronic RapID Compendium) Version 1.0.77, Catalog Number R8323600, used in conjunction with: remel...
Remel, Inc began a recall of ERIC (Electronic RapID Compendium) Version 1.0.77, Catalog Number R8323600, used in conjunction with: remel RapID Yeast Plus System is a qualitative... on . The reason FDA records is "There is an error in Version 1.0.77 of the electronic code compendium (ERIC) software. The error is limited to the database corresponding to the R8311007 RapID Yeast Plus System. The database erroneously includes...". As of , the recall status on record is Terminated, and MDM tags the root cause as software.
Recall record
- Recall number
- Z-0870-2010
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Remel Inc
- Product
- ERIC (Electronic RapID Compendium) Version 1.0.77, Catalog Number R8323600, used in conjunction with: remel RapID Yeast Plus System is a qualitative micromethod employing conventional and chromogenic substrates for the identification of medically important yeast, yeast-like, and related organisms isolated from human clinical specimens. A complete listing of the organisms addressed by the RapID Yeast Plus System is given in the RapID Yeast Plus Differential Chart. RapID" YEAST PLUS System, Catalog No. R8311007. 4 hour identification of yeast based on enzyme technology. Database includes over 40 taxa. Requires Inoculation Fluid, REF R8325106. CE, 20 Panels/Kit The responsible firm name on the product labeling is Remel, 12076 Santa Fe Dr., Lenexa, KS 66215-3594
- Product code
- JXB Kit, Identification, Yeast
- Reason for recall
- There is an error in Version 1.0.77 of the electronic code compendium (ERIC) software. The error is limited to the database corresponding to the R8311007 RapID Yeast Plus System. The database erroneously includes Candida dubliniensis as well as displays an incorrect temperature range as an accessory test for codes with an overlap between Candida albicans and Candida dubliniensis. When using the
- Root cause tag
- software
- FDA root cause
- Software design
- Action
- The firm issued an "IMPORTANT MEDICAL DEVICE PRODUCT RECALL NOTICE" dated October 29, 2009. After completion of the firm's investigation and risk assessment, Remel decided to notify customers of this situation and quarantine any remaining units on hand. The notification provided the consignees with instructions for obtaining and downloading the software patch. A response form was included with the recall letter and the consignee was asked to complete the form and return it to Remel. The consignees should contact the Technical Services Department at 800-255-6730 (USA) or 913-888-0939 (International), if they have inquiries concerning this advisory.
- Quantity in commerce
- 143 units of software
- Distribution
- Worldwide Distribution: USA, Europe and Canada
- Product codes and lots
- ERIC (Electronic RapID Compendium) Version 1.0.77 -- Lot No. 0001096277