Z-0667-2010

AC Power Cords, 110V Manufactured by Electri-Cord Manufacturing Company in Westfield, PA. Power Cord M426884...

KCI USA, Inc. / initiated 2009-11-04 / Terminated / root cause: manufacturing process

KCI USA, Inc. began a recall of AC Power Cords, 110V Manufactured by Electri-Cord Manufacturing Company in Westfield, PA. Power Cord M426884 used with VAC ATS Canadian Model M8259994 on . The reason FDA records is "Power cords may crack and fail inside plug with potential for fire hazard". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-0667-2010
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
KCI USA, Inc.
Product
AC Power Cords, 110V Manufactured by Electri-Cord Manufacturing Company in Westfield, PA. Power Cord M426884 used with VAC ATS Canadian Model M8259994.
Product code
OMP Negative Pressure Wound Therapy Powered Suction Pump
Reason for recall
Power cords may crack and fail inside plug with potential for fire hazard.
Root cause tag
manufacturing process
FDA root cause
Nonconforming Material/Component
Action
Kinetics Concepts, Inc. (KCI) issued an "Urgent Medical Device Correction" letter dated November 4, 2009. Consignees were informed of the affected product and instructed to take necessary action to avoid any further issues associated with the product. For further information, contact KCI's technical support team at 1-800-275-4524 option #3.
Quantity in commerce
649
Distribution
Nationwide and Canada
Product codes and lots
Device Model #M8259994 export only