Medical Device
Manufacturing
THE TRADE PAPER OF THE FACTORY FLOOR
Z-1042-2010

Calcium (CPC) R1 Reagent Set. Pointe Scientific Liquid Calcium (CPC) R1 Reagent Set, Pointe Scientific, Inc...

Horiba Instruments Incorporated / initiated 2009-11-10 / Terminated / root cause: manufacturing process

Pointe Scientific, Inc. began a recall of Calcium (CPC) R1 Reagent Set. Pointe Scientific Liquid Calcium (CPC) R1 Reagent Set, Pointe Scientific, Inc, Canton, MI; Catalog #'s HC402-R1. Calcium CPC... on . The reason FDA records is "Non-compliance with the Good Manufacturing Practice regulations, as the manufacturing process for this product had not been validated". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-1042-2010
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Horiba Instruments Incorporated
Product
Calcium (CPC) R1 Reagent Set. Pointe Scientific Liquid Calcium (CPC) R1 Reagent Set, Pointe Scientific, Inc, Canton, MI; Catalog #'s HC402-R1. Calcium CPC Hitachi Reagent Set for repackaging, Catalog # 3-HC902-R1. The product is used for the quantitative determination of calcium in serum on Hitachi analyzers.
Product code
CIC Cresolphthalein Complexone, Calcium
Reason for recall
Non-compliance with the Good Manufacturing Practice regulations, as the manufacturing process for this product had not been validated.
Root cause tag
manufacturing process
FDA root cause
Process control
Action
Consignees were notified by "Urgent: Medical Device Recall" letter dated November 11, 2009 and instructed to conduct a recall to the retail level, that users should dispose of the product immediately, and that a business reply card should be returned to account for products. If you have any questions, please call Pointe Scientific, Inc. Technical Support Department at 1-800-757-5313 or 1-734-487-8300.
Quantity in commerce
6 sets and 21 L
Distribution
Worldwide Distribution (United States, Chile, India, Republic of Yemen and Trinidad).
Product codes and lots
All lots containing 813401 as part of the lot number.