Calcium (CPC) R2 Reagent Set. Pointe Scientific Liquid Calcium (CPC) R2 Reagent Set, Pointe Scientific, Inc...
Pointe Scientific, Inc. began a recall of Calcium (CPC) R2 Reagent Set. Pointe Scientific Liquid Calcium (CPC) R2 Reagent Set, Pointe Scientific, Inc, Canton, MI; Catalog #'s HC402-R2. Calcium CPC... on . The reason FDA records is "Non-compliance with the Good Manufacturing Practice regulations, as the manufacturing process for this product had not been validated". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.
Recall record
- Recall number
- Z-1043-2010
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Horiba Instruments Incorporated
- Product
- Calcium (CPC) R2 Reagent Set. Pointe Scientific Liquid Calcium (CPC) R2 Reagent Set, Pointe Scientific, Inc, Canton, MI; Catalog #'s HC402-R2. Calcium CPC Hitachi Reagent Set for repackaging, Catalog # 3-HC902-R2. The product is used for the quantitative determination of calcium in serum on Hitachi analyzers.
- Product code
- CIC Cresolphthalein Complexone, Calcium
- Reason for recall
- Non-compliance with the Good Manufacturing Practice regulations, as the manufacturing process for this product had not been validated.
- Root cause tag
- manufacturing process
- FDA root cause
- Process control
- Action
- Consignees were notified by "Urgent: Medical Device Recall" letter dated November 11, 2009 and instructed to conduct a recall to the retail level, that users should dispose of the product immediately, and that a business reply card should be returned to account for products. If you have any questions, please call Pointe Scientific, Inc. Technical Support Department at 1-800-757-5313 or 1-734-487-8300.
- Quantity in commerce
- 4 sets and 9 L
- Distribution
- Worldwide Distribution (United States, Chile, India, Republic of Yemen and Trinidad).
- Product codes and lots
- All lots containing 813401 as part of the lot number.