Z-2335-2010

Abbott Spine PathFinder, Cannulated Poly Screwdriver II - MOD, non-sterile, Manufactured by Abbott Spine...

Zimmer GmbH / initiated 2009-01-12 / Terminated

Zimmer Inc. began a recall of Abbott Spine PathFinder, Cannulated Poly Screwdriver II - MOD, non-sterile, Manufactured by Abbott Spine, Inc., Austin, TX; Ref. 1160-2. Usage: The screwdriver... on . The reason FDA records is "The screwdriver may fracture, resulting in surgical delays and/or fragments going into the surgical site". As of , the recall status on record is Terminated.

Recall record

Recall number
Z-2335-2010
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Zimmer GmbH
Product
Abbott Spine PathFinder, Cannulated Poly Screwdriver II - MOD, non-sterile, Manufactured by Abbott Spine, Inc., Austin, TX; Ref. 1160-2. Usage: The screwdriver is intended to mate to the hexalobular feature of a PathFinder open polyaxial screw and threads onto the head to hold it in on-axis for implanting.
Product code
MNI Orthosis, Spinal Pedicle Fixation
Reason for recall
The screwdriver may fracture, resulting in surgical delays and/or fragments going into the surgical site.
FDA root cause
Other
Action
The sales force was provided with new instructions for use on 1/12/09 and again in 05/09. On April 6, 2010 the firm sent out an Urgent Medical Device Correction to it's field representatives for distribution to it's customers. The field representatives were to distribute the forms and provide a list of the customers that recieved the recall notice. For questions or assistance, please contact Zimmer Spine at 800-326-0635.
Quantity in commerce
691
Distribution
Worldwide.
Product codes and lots
All units. Lots 36WC, 39BA, 43MN, 44WP, 47FE, 18TH, 20XF, 25WE, 29CQ, 29YJ, 31TJ, 33ZH, 47RM, 48PC, 49QW, 51PF and 58SD.