Z-0499-2010
Zimmer Dental AdVent Implant System, Abutment, Tapered, Rx, sterile, Zimmer Dental, Inc., Carlsbad, CA; REF...
Zimmer GmbH / initiated 2009-11-06 / Terminated / root cause: manufacturing process
Zimmer Inc. began a recall of Zimmer Dental AdVent Implant System, Abutment, Tapered, Rx, sterile, Zimmer Dental, Inc., Carlsbad, CA; REF AVACT on . The reason FDA records is "The devices may not meet specifications, which may result in fracturing during tightening of the component". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.
Recall record
- Recall number
- Z-0499-2010
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Zimmer GmbH
- Product
- Zimmer Dental AdVent Implant System, Abutment, Tapered, Rx, sterile, Zimmer Dental, Inc., Carlsbad, CA; REF AVACT.
- Product code
- DZE Implant, Endosseous, Root-Form
- Reason for recall
- The devices may not meet specifications, which may result in fracturing during tightening of the component.
- Root cause tag
- manufacturing process
- FDA root cause
- Process control
- Action
- User accounts were notified of the issue by phone on or about 11/6/09. Clinicians were requested to schedule a follow-up appointment and to remove and replace the recalled components for any patients who had this device placed without fracturing during the procedure. Questions are directed to the company at 1-800-854-7019.
- Quantity in commerce
- 738
- Distribution
- Ohio, Canada and France.
- Product codes and lots
- Lots 61233652, 61238956, 61245049, 61252577, 61256798, 61263053, 61314067 and 61341224.
FIRST ARTICLE / WEEKLY