Z-1535-2010

Dimension Vista CHEM 3 calibrator is an in vitro diagnostic product for the calibration of ammonia (AMON)...

Siemens Healthcare Diagnostics Inc. / initiated 2009-10-23 / Terminated

Siemens Healthcare Diagnostics, Inc. began a recall of Dimension Vista CHEM 3 calibrator is an in vitro diagnostic product for the calibration of ammonia (AMON), Cardon Dioxide (CO2) and Ethyl Alcohol (ETOH) on the... on . The reason FDA records is "patient results are depressed". As of , the recall status on record is Terminated.

Recall record

Recall number
Z-1535-2010
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Siemens Healthcare Diagnostics Inc.
Product
Dimension Vista CHEM 3 calibrator is an in vitro diagnostic product for the calibration of ammonia (AMON), Cardon Dioxide (CO2) and Ethyl Alcohol (ETOH) on the Dimension Vista System
Product code
JIX Calibrator, Multi-Analyte Mixture
Reason for recall
patient results are depressed
FDA root cause
Other
Action
An "Urgent Field Safety Notice" dated October 2009 was issued by Siemens to all customers. The notice described the product, problem and actions to be taken by the customers. The customers were instructed to recalibrate the AMON method (AMON Flex reagent cartridge, K3019) with Vista Chem 3 Claibrator lot 9ED058 using the revised bottle values provided in the notice and to please consult with their laboratory director regarding re-evaluation of previous results. The customers should forward this notification to anyone to whom they may have distributed this product. If you have any technical questions regarding this information, please contact Siemens Technical Solutions Center at 1-800-441-9250.
Quantity in commerce
1753 cartons
Distribution
Worldwide distribution: USA and Australia, Canada, Japan, New Zealand, South Korea, and Switzerland.
Product codes and lots
Catalog number KC130. Lot number 9ED058 exp 5/1/10