Z-1535-2010
Dimension Vista CHEM 3 calibrator is an in vitro diagnostic product for the calibration of ammonia (AMON)...
Siemens Healthcare Diagnostics Inc. / initiated 2009-10-23 / Terminated
Siemens Healthcare Diagnostics, Inc. began a recall of Dimension Vista CHEM 3 calibrator is an in vitro diagnostic product for the calibration of ammonia (AMON), Cardon Dioxide (CO2) and Ethyl Alcohol (ETOH) on the... on . The reason FDA records is "patient results are depressed". As of , the recall status on record is Terminated.
Recall record
- Recall number
- Z-1535-2010
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Siemens Healthcare Diagnostics Inc.
- Product
- Dimension Vista CHEM 3 calibrator is an in vitro diagnostic product for the calibration of ammonia (AMON), Cardon Dioxide (CO2) and Ethyl Alcohol (ETOH) on the Dimension Vista System
- Product code
- JIX Calibrator, Multi-Analyte Mixture
- Reason for recall
- patient results are depressed
- FDA root cause
- Other
- Action
- An "Urgent Field Safety Notice" dated October 2009 was issued by Siemens to all customers. The notice described the product, problem and actions to be taken by the customers. The customers were instructed to recalibrate the AMON method (AMON Flex reagent cartridge, K3019) with Vista Chem 3 Claibrator lot 9ED058 using the revised bottle values provided in the notice and to please consult with their laboratory director regarding re-evaluation of previous results. The customers should forward this notification to anyone to whom they may have distributed this product. If you have any technical questions regarding this information, please contact Siemens Technical Solutions Center at 1-800-441-9250.
- Quantity in commerce
- 1753 cartons
- Distribution
- Worldwide distribution: USA and Australia, Canada, Japan, New Zealand, South Korea, and Switzerland.
- Product codes and lots
- Catalog number KC130. Lot number 9ED058 exp 5/1/10
FIRST ARTICLE / WEEKLY
What cleared, what was recalled, and what changed in the standards. One edition a week for the people who make medical devices.