Maquet Servo Ventilator 300/300A (SV300) Siemens - Elema AB
Maquet Cardiovascular Us Sales, LLC / initiated 2009-11-17 / Terminated / root cause: design
MAQUET Inc. began a recall of Maquet Servo Ventilator 300/300A (SV300) Siemens - Elema AB on . The reason FDA records is "Some Servo Ventilators 300/300A and Servo Ventilator 900C/D/E should not be used with a system that may generate negative pressure below -100 cm H2O (Closed system suctioning) due to the risk of damaging the...". As of , the recall status on record is Terminated, and MDM tags the root cause as design.
Recall record
- Recall number
- Z-0588-2010
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Maquet Cardiovascular Us Sales, LLC
- Product
- Maquet Servo Ventilator 300/300A (SV300) Siemens - Elema AB,
- Product code
- CBK Ventilator, Continuous, Facility Use
- Reason for recall
- Some Servo Ventilators 300/300A and Servo Ventilator 900C/D/E should not be used with a system that may generate negative pressure below -100 cm H2O (Closed system suctioning) due to the risk of damaging the ventilator's pressure transducers.
- Root cause tag
- design
- FDA root cause
- Device Design
- Action
- Maquet sent Device Correction letters commencing 11/17/09 and ending 11/23/09 by Federal Express.
- Quantity in commerce
- 20837 (7847 of those in US)
- Distribution
- Worldwide distribution
- Product codes and lots
- Serial numbers 00410-24999. Ventilators in the specific range that are marked with one of the following labels are already equipped with the new type ot pressure transducer. 1. Upgrade EM008/02/1; 2. Upgrade SV300 66 82 745 SV900 66 82 747