Maquet Servo Ventilator 300/300A (SV300) Siemens - Elema AB

Maquet Cardiovascular Us Sales, LLC / initiated 2009-11-17 / Terminated / root cause: design

MAQUET Inc. began a recall of Maquet Servo Ventilator 300/300A (SV300) Siemens - Elema AB on . The reason FDA records is "Some Servo Ventilators 300/300A and Servo Ventilator 900C/D/E should not be used with a system that may generate negative pressure below -100 cm H2O (Closed system suctioning) due to the risk of damaging the...". As of , the recall status on record is Terminated, and MDM tags the root cause as design.

Recall record

Recall number
Z-0588-2010
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Maquet Cardiovascular Us Sales, LLC
Product
Maquet Servo Ventilator 300/300A (SV300) Siemens - Elema AB,
Product code
CBK Ventilator, Continuous, Facility Use
Reason for recall
Some Servo Ventilators 300/300A and Servo Ventilator 900C/D/E should not be used with a system that may generate negative pressure below -100 cm H2O (Closed system suctioning) due to the risk of damaging the ventilator's pressure transducers.
Root cause tag
design
FDA root cause
Device Design
Action
Maquet sent Device Correction letters commencing 11/17/09 and ending 11/23/09 by Federal Express.
Quantity in commerce
20837 (7847 of those in US)
Distribution
Worldwide distribution
Product codes and lots
Serial numbers 00410-24999. Ventilators in the specific range that are marked with one of the following labels are already equipped with the new type ot pressure transducer. 1. Upgrade EM008/02/1; 2. Upgrade SV300 66 82 745  SV900 66 82 747