Z-0589-2010
Maquet Servo Ventilator SV900C/D/E (SV900) Siemens-Elema AB, Solna, Sweden
Maquet Cardiovascular Us Sales, LLC / initiated 2009-11-17 / Terminated / root cause: design
MAQUET Inc. began a recall of Maquet Servo Ventilator SV900C/D/E (SV900) Siemens-Elema AB, Solna, Sweden on . The reason FDA records is "Some Servo Ventilators 300/300A and Servo Ventilator 900C/D/E should not be used with a system that may generate negative pressure below -100 cm H2O (Closed system suctioning) due to the risk of damaging the...". As of , the recall status on record is Terminated, and MDM tags the root cause as design.
Recall record
- Recall number
- Z-0589-2010
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Maquet Cardiovascular Us Sales, LLC
- Product
- Maquet Servo Ventilator SV900C/D/E (SV900) Siemens-Elema AB, Solna, Sweden
- Product code
- CBK Ventilator, Continuous, Facility Use
- Reason for recall
- Some Servo Ventilators 300/300A and Servo Ventilator 900C/D/E should not be used with a system that may generate negative pressure below -100 cm H2O (Closed system suctioning) due to the risk of damaging the ventilator's pressure transducers.
- Root cause tag
- design
- FDA root cause
- Device Design
- Action
- Maquet sent Device Correction letters commencing 11/17/09 and ending 11/23/09 by Federal Express.
- Quantity in commerce
- 29364 (8438 of those in US)
- Distribution
- Worldwide distribution
- Product codes and lots
- Serial Numbers 010001-188499 Ventilators within the specified serial number range, that are marked with one of the following labels on the inside of the lid of the pneumatic unit are exempted since they have already been fitted with a new type of the pressure transducer. 1. Upgrade EMO 12/02/1; 2. Upgrade SV300 6682-745 SV900 66 82 747.