Stryker Orthopaedics Sterile Knee Instrument, Patella Cutter; Stainless Steel, Disposable; Howmedica...

Howmedica Osteonics Corp / initiated 2009-11-11 / Terminated / root cause: labeling and UDI

Stryker Howmedica Osteonics Corp. began a recall of Stryker Orthopaedics Sterile Knee Instrument, Patella Cutter; Stainless Steel, Disposable; Howmedica Osteonics Corp., 325 Corporate Drive, Mahwah, NJ 07430 33... on . The reason FDA records is "Disposable Patella Cutters - 30 mm (catalog number 8000-0030, lot V31257001) were packaged and labeled as 33 mm Disposable Patella Cutters (catalog number 8000-0033, lot V31257001 and vice versa". As of , the recall status on record is Terminated, and MDM tags the root cause as labeling and UDI.

Recall record

Recall number
Z-0529-2010
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Howmedica Osteonics Corp
Product
Stryker Orthopaedics Sterile Knee Instrument, Patella Cutter; Stainless Steel, Disposable; Howmedica Osteonics Corp., 325 Corporate Drive, Mahwah, NJ 07430 33 mm size, One single unit to a package. Note: The 30 mm instruments were not distributed. The patella cutters are components of the Xcelerate Instruments and are used with additional devices to remove bone from the patella in preparation for a replacement implant. The Xcelerate Patella Reaming Instruments provide a means to prepare for implantation of either the Duracon or Scorpio Total Knee Systems patellar implants.
Product code
HTS Knife, Orthopedic
Reason for recall
Disposable Patella Cutters - 30 mm (catalog number 8000-0030, lot V31257001) were packaged and labeled as 33 mm Disposable Patella Cutters (catalog number 8000-0033, lot V31257001 and vice versa.
Root cause tag
labeling and UDI
FDA root cause
Labeling mix-ups
Action
An "Urgent Product Recall" letter dated November 11, 2009 was sent to all affected Stryker branches, hospital risk management, hospital Chief or Orthopaedics and surgeons by Federal Express. The letter described the issue, potential hazards and risk mitigation factors and requested that the Recall Acknowledgement form be completed and returned. The Recall Acknowledgement form has information regarding the return of affected product to the firm. Direct questions to Stryker Orthopaedics by calling 1-201-831-5832.
Quantity in commerce
12
Distribution
Nationwide Distribution (MA, NJ, FL, TN, OR and NY).
Product codes and lots
Catalog Number 8000-0033; Lot code V31257001.