PLAC Test Reagent Kit, immuno-turbidimetric method, Part Number 10-112, 10-0113, and 10-115 Advia test kit...
DiaDexus, Inc began a recall of PLAC Test Reagent Kit, immuno-turbidimetric method, Part Number 10-112, 10-0113, and 10-115 Advia test kit, manufactured by DiaDexus, South San Francisco, CA on . The reason FDA records is "Inaccurate results-- some samples may return with falsely elevated results potentially as a result of an interfering substance. Some patients may be subjected to additional lifestyle changes or intensified...". As of , the recall status on record is Terminated.
Recall record
- Recall number
- Z-1753-2010
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Diadexus, Inc.
- Product
- PLAC Test Reagent Kit, immuno-turbidimetric method, Part Number 10-112, 10-0113, and 10-115 Advia test kit, manufactured by DiaDexus, South San Francisco, CA.
- Product code
- NOE Test, System, Immunoassay, Lipoprotein-Associated Phospholipase A2
- Reason for recall
- Inaccurate results-- some samples may return with falsely elevated results potentially as a result of an interfering substance. Some patients may be subjected to additional lifestyle changes or intensified pharmaceutical therapy; such as statins, as a result of the falsely elevated result.
- FDA root cause
- Other
- Action
- Notification letter issued to consignees on 10/29/2009. Customers were instructed to not report PLAC test results for samples that are greater than 360 ng/mL, and pursue confirmatory testing by contacting DiaDexus. Customers with questions please call the company technical support line toll free at 1-877-752-2837 for assistance. UPDATE: On April 30, the firm suspended marketing of the PLAC TIA method, as the limited range no longer covered the clinical data/reference range data cited in the product insert. The firm sent out a letter notifying customers on May 10, 2010. The firm set up a sample send out program, so that customers can send samples to labs running the ELISA method PLAC test. Arrangements were made so that labs could be enabled to run a CLIA complex assay, for customer credit programs for un-used products, and a fax back response form for response tracking.
- Quantity in commerce
- all lots
- Distribution
- Product was distributed to 123 consignees throughout the US.
- Product codes and lots
- Lot numbers 078011, 088061, 098091, 109011, 119021, 129031, 139051, 139051A