Medical Device
Manufacturing
THE TRADE PAPER OF THE FACTORY FLOOR
Z-0478-2010

Protg EverFlex" Self-Expanding Biliary Stent System (for United States distribution). 6 x 150mm, Model #...

Medtronic, Inc. / initiated 2009-11-06 / Terminated / root cause: labeling and UDI

Ev3, Inc. began a recall of Protg EverFlex" Self-Expanding Biliary Stent System (for United States distribution). 6 x 150mm, Model # PRB35-06-150-080. Sterile EO, ev3 Inc, 4600 Nathan... on . The reason FDA records is "ev3 has determined that a lot of 150mm Protg EverFlex Biliary Stent System contains a 100mm length Self-Expanding stent. The implantation of a stent 50mm shorter than expected may not cover the full length of the target...". As of , the recall status on record is Terminated, and MDM tags the root cause as labeling and UDI.

Recall record

Recall number
Z-0478-2010
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Medtronic, Inc.
Product
Protg EverFlex" Self-Expanding Biliary Stent System (for United States distribution). 6 x 150mm, Model # PRB35-06-150-080. Sterile EO, ev3 Inc, 4600 Nathan lane North, Plymouth, MN 55442-2920, Rx only. Protg EverFlex" Self-Expanding Peripheral Stent System (for Outside United States distribution). 6 x 100mm, Model Number PRP35-06-100-080. Sterile EO, ev3 Inc, 4600 Nathan lane North, Plymouth, MN 55442-2920, Rx only. In the United States, the stent is intended as a palliative treatment of malignant neoplasms in the biliary tree. Outside the United States, it is intended indicated for use in occlusions, lesions at high risk for abrupt closure or threatened closure following percutaneous transluminal angioplasty (PTA) or lesions believed to be at high risk for restenosis following PTA in the common and external iliac, superficial femoral, proximal popliteal, or subclavian arteries.
Product code
FGE Stents, Drains And Dilators For The Biliary Ducts
Reason for recall
ev3 has determined that a lot of 150mm Protg EverFlex Biliary Stent System contains a 100mm length Self-Expanding stent. The implantation of a stent 50mm shorter than expected may not cover the full length of the target lesion and may require subsequent intervention.
Root cause tag
labeling and UDI
FDA root cause
Labeling mix-ups
Action
Two separate letters were sent to United States and international consignees on November 9, 2009. The letter to US consignees was addressed to Cath Lab Managers while the letter to foreign consignees was addressed to the Risk Manager or Cath Lab Manager. The letters described the affected product, gave an Issue Summary and Required Action, which advised consignees to locate and remove the recalled product. An Ev3, Inc. sales representative will contact consignees to arrange return of all unused product to Ev3, Inc. Direct questions regarding the recall to Ev3, Inc. by calling Customer Service at 1-800-716-6700.
Quantity in commerce
6 (United States) and 6 (Outside United States)
Distribution
Worldwide Distribution -- United States (TX and FL), GERMANY, FRANCE and UNITED KINGDOM.
Product codes and lots
United States lot # 7821740 and  Outside United States lot # 7821736