Invacare Storm TDX SR Power Wheelchair with Stability Lock Model numbers: TDXSR, TDXSR-CG, TDXSR-CG-HD...

Invacare GmbH / initiated 2009-09-15 / Terminated / root cause: design

Invacare Corporation began a recall of Invacare Storm TDX SR Power Wheelchair with Stability Lock Model numbers: TDXSR, TDXSR-CG, TDXSR-CG-HD, TDXSR-HD, TDXSR-MCG, TDXSR-MCG-HD, TDXSRV-HD on . The reason FDA records is "For some Invacare chairs, the stability lock feature may not be engaging properly or consistently. If the stability lock feature does not engage properly, the wheelchair may do one or more of the following: (1) veer to...". As of , the recall status on record is Terminated, and MDM tags the root cause as design.

Recall record

Recall number
Z-0579-2010
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Invacare GmbH
Product
Invacare Storm TDX SR Power Wheelchair with Stability Lock Model numbers: TDXSR, TDXSR-CG, TDXSR-CG-HD, TDXSR-HD, TDXSR-MCG, TDXSR-MCG-HD, TDXSRV-HD
Product code
ITI Wheelchair, Powered
Reason for recall
For some Invacare chairs, the stability lock feature may not be engaging properly or consistently. If the stability lock feature does not engage properly, the wheelchair may do one or more of the following: (1) veer to one side; (2) rock forward onto its front riggings; (3) drive in an unintended circular motion; and/or (4) fall forward or to one side. In these cases there is a risk to injury for
Root cause tag
design
FDA root cause
Device Design
Action
On 9/15/2009 the firm sent recall notification letter to their customers instructing them to immediately contact the dealer to have the stability lock inspected and/or adjusted. If the dealer is unavailable or the repair will take longer than 60 days, they are to contact the firm and arrange for an inspection and/or adjustment. Invacare requested contact names to whom it can send additional information. Invacare plans to mail, via FEDEX, information regarding the recall to each consignee. Initial consignees received a response card and instructions to forward cards to any customer to whom they have distributed the product. The products are to be inspected at the cylinder label and if the label does not have a long alpha-numeric string demonstrated in the label, the cylinders on the chair have to be replaced. Questions are directed to [email protected] or to 800-333-6900.
Quantity in commerce
36
Distribution
Product was distributed to the following states: AZ, CA, CO, CT, DC, FL, GA, IA, IL, IN, KS, MA, ME, MI, MN, MO, MS, NC, ND, NM, NY, OH, SC & TX. Product was also distributed in Canada.
Product codes and lots
Model numbers: TDXSR, TDXSR-CG, TDXSR-CG-HD, TDXSR-HD, TDXSR-MCG, TDXSR-MCG-HD, TDXSRV-HD. Serial Numbers: 08BE000082; 08BE000088; 08BE000089; 08BE000090; 08BE000323; 08BE001102; 08BE001105; 08BE002283; 08BE002803; 08BE003025; 08BE003250; 08BE003251; 08BE003472; 08BE003473; 08BE003632; 08BE004402; 08BE004414; 08BE004415; 08BE004835; 08BE005297; 08DE002040; 08EE000277; 08EE001539; 08EE001540; 08EE001775; 08FE003254; 08FE003257; 08FE003668; 08FE004681; 08GE003570; 08GE003936; 08HE001169; 08HE005013; 08IE001081; 08IE001316 & 08IE002729.