TrailBlazer Support Catheter, ev3 inc, 4600 Nathan Lane North, Plymouth, MN 55442-2920, Rx Only, Sterile EO...
Ev3, Inc. began a recall of TrailBlazer Support Catheter, ev3 inc, 4600 Nathan Lane North, Plymouth, MN 55442-2920, Rx Only, Sterile EO. US Model Numbers: SC-014-135, SC-018-090... on . The reason FDA records is "Catheter may crack near the radiopaque marker band at the distal end of the device during use. Cracking of the catheter shaft in this location can result in material separation, and potential embolization". As of , the recall status on record is Terminated, and MDM tags the root cause as design.
Recall record
- Recall number
- Z-0497-2010
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Medtronic, Inc.
- Product
- TrailBlazer Support Catheter, ev3 inc, 4600 Nathan Lane North, Plymouth, MN 55442-2920, Rx Only, Sterile EO. US Model Numbers: SC-014-135, SC-018-090, SC-035-065, SC-035-135, SC-014-150, SC-018-150, SC-035-090, and SC-035-150. Intended to guide and support a guide wire during access of the vasculature.
- Product code
- DQY Catheter, Percutaneous
- Reason for recall
- Catheter may crack near the radiopaque marker band at the distal end of the device during use. Cracking of the catheter shaft in this location can result in material separation, and potential embolization.
- Root cause tag
- design
- FDA root cause
- Component design/selection
- Action
- Consignees were sent a Medical Device Recall letter, dated 11/6/09. The letter was addressed to Risk Manger or Cath Lab Manage. The letter provided the Issue Summary, identified the Affected Product, and stated the Required Action, which asked consignees to locate and remove the affected product. Questions should be directed to ev3 Customer Service at 800-716-6700.
- Quantity in commerce
- 70
- Distribution
- Nationwide Distribution -- USA, including states of CA, FL, GA, IL, IN, IA, KS, KY, MI, MT, NJ, PA, SD, TN, TX, and WI.
- Product codes and lots
- Lot #: 7828282, 7792290, 7792584, 7805570, 7805797, 7806392, 7820252, 7790666, 7791887, 7803307, 7835331, 7822400, 7822593, 7800446, 7800555, 7800756, 7800779, 7800809, 7801822, 7801875, 7803305, 7803306, 7820273, 7834779, 7834845, 7824905, and 7832205.