Z-1073-2010
Medical Action Industries, IV Start Bundle Kit, using MPS Acacia with BD Q-Syte (BD #385151), Part number...
Medical Action Industries, Inc. 306 / initiated 2009-11-06 / Terminated / root cause: manufacturing process
Medical Action Industries Inc began a recall of Medical Action Industries, IV Start Bundle Kit, using MPS Acacia with BD Q-Syte (BD #385151), Part number: 69946B on . The reason FDA records is "Manufacturing defect may allow air entry into a central venous catheter, possibly resulting in an air embolism". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.
Recall record
- Recall number
- Z-1073-2010
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Medical Action Industries, Inc. 306
- Product
- Medical Action Industries, IV Start Bundle Kit, using MPS Acacia with BD Q-Syte (BD #385151), Part number: 69946B.
- Product code
- FPA Set, Administration, Intravascular
- Reason for recall
- Manufacturing defect may allow air entry into a central venous catheter, possibly resulting in an air embolism.
- Root cause tag
- manufacturing process
- FDA root cause
- Nonconforming Material/Component
- Action
- On 11/6/2009 a letter was drafted and sent by email to three (3) Medical Action Industries customers. They were instructed to return the recalled products in inventory to MAI. They further advised, if the product was distributed to notify their customer, inform them of the recall and instruct them to return the affected lots to MAI.
- Quantity in commerce
- 12,000 kits
- Distribution
- Nationwide distribution
- Product codes and lots
- Part number: 69946B, Lot numbers: 62295, 63591, 64197, 65175, 67162, 69906, 72113 and 73617.