Gyrus ACMI PKS Seal Open Forceps, model 915005PK, 9-3/4" Straight Jaw. Sterile EO. The device is intended to...

Gyrus ACMI, Inc. / initiated 2009-08-18 / Terminated / root cause: manufacturing process

Gyrus ACMI Corporation began a recall of Gyrus ACMI PKS Seal Open Forceps, model 915005PK, 9-3/4" Straight Jaw. Sterile EO. The device is intended to be used with the G400 electrosurgical generator... on . The reason FDA records is "Gyrus ACMI received complaints of a problem with 3103PK, 3104PK, 3105PK, 915000PK, 915005PK, 915010PK - Gyrus ACMI PKS Seal Open Forceps. The metal shim between the jaws may detach during the procedure and fall into the...". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-0636-2010
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Gyrus ACMI, Inc.
Product
Gyrus ACMI PKS Seal Open Forceps, model 915005PK, 9-3/4" Straight Jaw. Sterile EO. The device is intended to be used with the G400 electrosurgical generator. Gyrus ACMI, Inc., 136 Turnpike Road, Southborough, MA 01772-2104 USA.
Product code
GEI Electrosurgical, Cutting & Coagulation & Accessories
Reason for recall
Gyrus ACMI received complaints of a problem with 3103PK, 3104PK, 3105PK, 915000PK, 915005PK, 915010PK - Gyrus ACMI PKS Seal Open Forceps. The metal shim between the jaws may detach during the procedure and fall into the surgical field. The detached shim may be undetected and be left behind in the patient. This circumstance could subsequently cause complications. Further use of this product should cease, affected product should be removed from stockrooms and returned to Gyrus ACMI.
Root cause tag
manufacturing process
FDA root cause
Environmental control
Action
An Urgent Medical Device Recall letter, dated 08/15/09 was sent to customers notifying them of the issue. The letter identified affected product, asked that they cease further use, remove affected devices from stockrooms, contact Gyrus ACMI customer service for a return goods authorization and return product to Gyrus ACMI at no charge. A reply form is also asked to be returned to Gyrus ACMI via fax. Questions and requests for a Return Goods Authorization are directed to Gyrus ACMI customer service at 1-888-524-7266.
Quantity in commerce
93,665 total (63,418 US, 30,247 OUS)
Distribution
US and International distribution. OUS to Include: Australia, Belgium, Bolivia, Bulgaria, Canada, Chile, China, Columbia, Croatia, Czech, Denmark, Ecuador, Estonia, Finland, France, Germany, Greece , Hong Kong, India, Ireland, Israel, Italy, Japan, Korea, Malaysia, Mexico, Netherlands, Panama, Puerto Rico, Singapore, Slovenia, South Africa, Spain, Sweden, Switzerland, Taiwan, United Kingdom.
Product codes and lots
all lots