Z-0615-2010

Haematokrit 20 Centrifuge with 2050 Rotor TYP #2014-01 and TYP#201424-01. Used for the purpose of determining...

Hettich Instruments LP / initiated 2009-11-21 / Terminated / root cause: design

Hettich Instruments Management I began a recall of Haematokrit 20 Centrifuge with 2050 Rotor TYP #2014-01 and TYP#201424-01. Used for the purpose of determining the hematocrit count of a patient sample by way... on . The reason FDA records is "Centrifuge Rotor Failure: may crack, break, and explode outside the centrifuge". As of , the recall status on record is Terminated, and MDM tags the root cause as design.

Recall record

Recall number
Z-0615-2010
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Hettich Instruments LP
Product
Haematokrit 20 Centrifuge with 2050 Rotor TYP #2014-01 and TYP#201424-01. Used for the purpose of determining the hematocrit count of a patient sample by way of applying centrifugal force in a manner that separates red blood cells and allows the proportion of red cell mass to plasma to be measured.
Product code
GKG Centrifuge, Hematocrit
Reason for recall
Centrifuge Rotor Failure: may crack, break, and explode outside the centrifuge.
Root cause tag
design
FDA root cause
Device Design
Action
Hettich Instruments notified Distributors by email on 11/21/09, to the Request For Urgent Action notifications, dated 11/23/2009. There is a description of how to identify a recalled rotor as well as a description of the replacement metal rotor that is to be used in place of the list of centrifuges affected by this recall. Customers are to cease using the setup described and contact their distributor or the firm. The firm will replace the plastic housing with a metal housing. Questions should be directed to 978-232-3957.
Quantity in commerce
46 units total
Distribution
Worldwide Distribution -- USA, including states of CT and IN, and country of Canada.
Product codes and lots
Units with plastic motor housing. TYP #2014-01 and TYP#201424-01.