Z-0781-2010

FilteredFlo Cardiac Blanket. The device is used to distribute heated air around a patient (for patient...

Gentherm Medical, LLC / initiated 2009-11-14 / Terminated / root cause: labeling and UDI

Cincinnati Sub-Zero Products Inc began a recall of FilteredFlo Cardiac Blanket. The device is used to distribute heated air around a patient (for patient warming) on . The reason FDA records is "Product is labeled as "Sterile". It was observed that incomplete seals were created in the packaging materials during production. When the seal is compromised, the sterility of the product inside the package cannot be...". As of , the recall status on record is Terminated, and MDM tags the root cause as labeling and UDI.

Recall record

Recall number
Z-0781-2010
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Gentherm Medical, LLC
Product
FilteredFlo Cardiac Blanket. The device is used to distribute heated air around a patient (for patient warming).
Product code
DWJ System, Thermal Regulating
Reason for recall
Product is labeled as "Sterile". It was observed that incomplete seals were created in the packaging materials during production. When the seal is compromised, the sterility of the product inside the package cannot be assured.
Root cause tag
labeling and UDI
FDA root cause
Other
Action
On November 14, 2009, Cincinnati Sub-zero Products, Inc. (CSZ) sent a recall notification letter and response letter to customers. A second notice dated December 10, 2009 was also issued to customers. Customers were instructed in the notifications to assume the recalled product is not sterile and do not use. CSZ requested all affected product be returned to their firm for replacement. Customers are to complete and return the response letter acknowledging receipt of the recall letter. Direct questions to CSZ by calling Customer Service at 1-800- 989-7373. Customers may also call 1-800-989-7373, extension 3234 or 1-51 3-71 9-3234 for recall instructions. Representatives will be available to take your call from 8:30 a.m. until 4:30 p.m. eastern standard time, Monday through Friday.
Quantity in commerce
1,095 Blankets
Distribution
Worldwide Distribution -- United States (AL, AR, AZ, CA, CO, CT, FL, GA, HI, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MO, MT, NC, ND, NE, NJ, NM, NY, OH, OK, OR, PA, RI, SC, TN, TX, UT, VA, WA, WI & WV), Canada, France, Germany, Ireland, Israel, New Zealand, Singapore, and The Netherlands.
Product codes and lots
Product # 82542; Catalog # 542; Lot #: 07279, 07280, 07424 & 08813.