Cordis Aquatrack Hydrophilic Nitinol Guidewire, for use in angiographic procedures, REF C3526-SSA, Guidewire...
Cordis Corporation began a recall of Cordis Aquatrack Hydrophilic Nitinol Guidewire, for use in angiographic procedures, REF C3526-SSA, Guidewire Diameter .035" Tip Length Flexible 3 cm, Sterile... on . The reason FDA records is "Some of the guidewires in these two lots have a straight tip (incorrect) rather than an angled tip (correct configuration for that catalog number)". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.
Recall record
- Recall number
- Z-0623-2010
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Cordis US Corp.
- Product
- Cordis Aquatrack Hydrophilic Nitinol Guidewire, for use in angiographic procedures, REF C3526-SSA, Guidewire Diameter .035" Tip Length Flexible 3 cm, Sterile. Contents 5 Hydrophilic guidewires and torquing devices.
- Product code
- DQX Wire, Guide, Catheter
- Reason for recall
- Some of the guidewires in these two lots have a straight tip (incorrect) rather than an angled tip (correct configuration for that catalog number).
- Root cause tag
- manufacturing process
- FDA root cause
- Process control
- Action
- Recall for Lot 80000521 was initiated on August 25, 2009 with Cordis Medical Device Recall letters being mailed for delivery on August 25, 2009. Response form was also provided to obtain product disposition. Recall for Lot 80000584 was initiated on October 19, 2009 with Cordis Medical Device Recall letters being mailed for delivery on October 20, 2009. Response form was also provided to obtain product disposition.
- Quantity in commerce
- 145 units
- Distribution
- Nationwide distribution.
- Product codes and lots
- Lots 80000521 and 80000584