TempPad Cold Therapy Pad. The device is used for localized temperature therapy including hand, elbow, arm...
Cincinnati Sub-Zero Products Inc began a recall of TempPad Cold Therapy Pad. The device is used for localized temperature therapy including hand, elbow, arm, foot, ankle, leg, or knee on . The reason FDA records is "Product is labeled as "Sterile". It was observed that incomplete seals were created in the packaging materials during production. When the seal is compromised, the sterility of the product inside the package cannot be...". As of , the recall status on record is Terminated, and MDM tags the root cause as labeling and UDI.
Recall record
- Recall number
- Z-0782-2010
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Gentherm Medical, LLC
- Product
- TempPad Cold Therapy Pad. The device is used for localized temperature therapy including hand, elbow, arm, foot, ankle, leg, or knee.
- Product code
- ILO Pack, Hot Or Cold, Water Circulating
- Reason for recall
- Product is labeled as "Sterile". It was observed that incomplete seals were created in the packaging materials during production. When the seal is compromised, the sterility of the product inside the package cannot be assured.
- Root cause tag
- labeling and UDI
- FDA root cause
- Other
- Action
- On November 14, 2009, Cincinnati Sub-zero Products, Inc. (CSZ) sent a recall notification letter and response letter to customers. A second notice dated December 10, 2009 was also issued to customers. Customers were instructed in the notifications to assume the recalled product is not sterile and do not use. CSZ requested all affected product be returned to their firm for replacement. Customers are to complete and return the response letter acknowledging receipt of the recall letter. Direct questions to CSZ by calling Customer Service at 1-800- 989-7373. Customers may also call 1-800-989-7373, extension 3234 or 1-51 3-71 9-3234 for recall instructions. Representatives will be available to take your call from 8:30 a.m. until 4:30 p.m. eastern standard time, Monday through Friday.
- Quantity in commerce
- 8,872 cartons of ten + 677 each (89,397 total)
- Distribution
- Worldwide Distribution -- United States (AL, AR, AZ, CA, CO, CT, FL, GA, HI, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MO, MT, NC, ND, NE, NJ, NM, NY, OH, OK, OR, PA, RI, SC, TN, TX, UT, VA, WA, WI & WV), Canada, France, Germany, Ireland, Israel, New Zealand, Singapore, and The Netherlands.
- Product codes and lots
- Part #: 50625, Catalog number: 510-US, Lot #: 08296, 08529, 08530, 08531, 08674, 08759, 08826, 08952, 09052, 09166, 09422, 09423; Part #: 50633, Catalog #: 510-USS, Lot #: 08537, 08675, 08953; Part #: 50627, Catalog #: 814-US, Lot #: 08610; Part # 50634, Catalog # 814-USS, Lot #: 08611, 08919, 09466; Part #: 50628, Catalog #: 1112-US, Lot #'s: 08382, 08411, 08475, 08476, 08477, 08557, 08583, 08657, 08666, 08667, 09486, 08668, 08707, 08771, 08810, 08874, 08903, 08971, 09037, 09088, 09089, 09149, 09225, 09251, 09285, 09313, 09314, 09327, 09360, 09361, 08320; Part #:50635, Catalog #: 1112-USS, Lot #'s: 08410, 08514, 08606, 08673, 08708, 08772, 08802, 08868, 08954, 09017, 09066, 09067, 09195, 09253, 09254, 09295, 09377, 09378, 09485, 08331; Part #: 50629, Catalog #: 1215-US, Lot #'s: 08395, 08584, 08669, 08732, 08917, 08989, 09107, 09218 & Part #: 50137, and Catalog #: CT-99, Lot #'s: 08332, 08353, 08438, 08538, 08607, 08608, 08706, 08725, 08773, 08837, 08851, 08879, 08933, 08970, 09010, 09086, 09087, 09143, 09144, 09157, 09294, 09329, 09362, 09379 and 08423.
FIRST ARTICLE / WEEKLY
What cleared, what was recalled, and what changed in the standards. One edition a week for the people who make medical devices.