Z-0598-2010

Stryker MIS Bur, 3.5mm Neuro Diamond, sterile, Stryker Instruments, Kalamazoo, MI; REF 8420-107-135

Stryker Corp. / initiated 2009-07-06 / Terminated / root cause: design

Stryker Instruments Div. of Stryker Corporation began a recall of Stryker MIS Bur, 3.5mm Neuro Diamond, sterile, Stryker Instruments, Kalamazoo, MI; REF 8420-107-135 on . The reason FDA records is "The bur may fracture, resulting in fragments within the surgical site, injury to the patient or surgical team, or both". As of , the recall status on record is Terminated, and MDM tags the root cause as design.

Recall record

Recall number
Z-0598-2010
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Stryker Corp.
Product
Stryker MIS Bur, 3.5mm Neuro Diamond, sterile, Stryker Instruments, Kalamazoo, MI; REF 8420-107-135.
Product code
HBE Drills, Burrs, Trephines & Accessories (Simple, Powered)
Reason for recall
The bur may fracture, resulting in fragments within the surgical site, injury to the patient or surgical team, or both.
Root cause tag
design
FDA root cause
Device Design
Action
Consignees were sent a letter dated 7/10/09 instructing them to locate and return product on hand for replacement. Surgeons who use the product were notified via a separate, untitled letter, dated July 13, 2009. A speed warning will be added to the product labels. Questions are directed to the firm at: 1-800-800-4236 ext. 3808.
Quantity in commerce
5438 of all products
Distribution
Nationwide distribution: Australia, Canada, England, France, Hong Kong, Japan, Netherlands, South Africa, Sweden, Switzerland and Taiwan.
Product codes and lots
All lots distributed 1/30/06 through 3/4/09. Lots 05363017, 06037017, 06069027, 06094017, 06165017, 06240017, 06275017, 06305017, 06354017, 07029017, 07079017, 07080017, 07108017, 07143017, 07219017, 07233017, 07290017, 07333017, 08021017, 08038017, 08049017, 08098017, 08155017, 08156017, 08253017, 08337017 and 09054017.