Z-0599-2010
Stryker MIS Bur, 3.0mm Neuro Diamond Course, sterile, Stryker Instruments, Kalamazoo, MI; REF 8420-107-230
Stryker Corp. / initiated 2009-07-06 / Terminated / root cause: design
Stryker Instruments Div. of Stryker Corporation began a recall of Stryker MIS Bur, 3.0mm Neuro Diamond Course, sterile, Stryker Instruments, Kalamazoo, MI; REF 8420-107-230 on . The reason FDA records is "The bur may fracture, resulting in fragments within the surgical site, injury to the patient or surgical team, or both". As of , the recall status on record is Terminated, and MDM tags the root cause as design.
Recall record
- Recall number
- Z-0599-2010
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Stryker Corp.
- Product
- Stryker MIS Bur, 3.0mm Neuro Diamond Course, sterile, Stryker Instruments, Kalamazoo, MI; REF 8420-107-230.
- Product code
- HBE Drills, Burrs, Trephines & Accessories (Simple, Powered)
- Reason for recall
- The bur may fracture, resulting in fragments within the surgical site, injury to the patient or surgical team, or both.
- Root cause tag
- design
- FDA root cause
- Device Design
- Action
- Consignees were sent a letter dated 7/10/09 instructing them to locate and return product on hand for replacement. Surgeons who use the product were notified via a separate, untitled letter, dated July 13, 2009. A speed warning will be added to the product labels. Questions are directed to the firm at: 1-800-800-4236 ext. 3808.
- Quantity in commerce
- 5438 of all products
- Distribution
- Nationwide distribution: Australia, Canada, England, France, Hong Kong, Japan, Netherlands, South Africa, Sweden, Switzerland and Taiwan.
- Product codes and lots
- All lots distributed 1/30/06 through 3/4/09. Lots 07352017, 08043017 and 08350027.