Z-0780-2010

MRI Eclipse 1.5T System Model #781230. The 1.5T Magnetic Resonance Imaging system intended for use with...

Philips North America LLC / initiated 2009-12-07 / Terminated

Philips Medical Systems (Cleveland) Inc began a recall of MRI Eclipse 1.5T System Model #781230. The 1.5T Magnetic Resonance Imaging system intended for use with clinical/radiologic diagnostic imaging applications on . The reason FDA records is "Polymeric RF connector block on the tabletop overheated and eventually ignited during scanning due to combination of arcing of electrical components associated with the RF trap assembly located below the connector and...". As of , the recall status on record is Terminated.

Recall record

Recall number
Z-0780-2010
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Philips North America LLC
Product
MRI Eclipse 1.5T System Model #781230. The 1.5T Magnetic Resonance Imaging system intended for use with clinical/radiologic diagnostic imaging applications.
Product code
LNH System, Nuclear Magnetic Resonance Imaging
Reason for recall
Polymeric RF connector block on the tabletop overheated and eventually ignited during scanning due to combination of arcing of electrical components associated with the RF trap assembly located below the connector and flame rating of the connector block material.
FDA root cause
Other
Action
An "Urgent-Field Safety Notice" dated December 4, 2009 was issued via certified mail to customers. The notification letter described the affected product, problem, hazard involved and action to be taken by customer. The customer will ensure to maintain frequent contact with patients during scanning and that the patient comfort kit is installed for all patient studies. The Customer will be visited by a Philips Service Representative to inspect their MRI System and install the newly remanufactured RF Connector Block. Additionally, step-by-step trap/cable installation instructions including pictures are being drafted to ensure correct installation of cabling. If you need any further information or support concerning this issue, please contact your local Philips representative or 1-800-722-9377.
Quantity in commerce
290 units
Distribution
Worldwide distribution- United states: (AL, AZ, CA, CO, CT, DC, FL, GA, IA, IL, IN, KS, KY, LA, LA, MA, MD, ME, MN, MO, NE, NH, NJ, NY, OH, OK, PA, RI, SC, TN, TX, UT, VA & WI), ARGENTINA, BRAZIL, CANADA, CHINA, CYPRUS, FRANCE, GERMANY, GREECE, ITALY, JAPAN, KOREA, NEW ZEALAND, PAKISTAN, POLAND, RUSSIA, SWITZERLAND, TAIWAN & UNITED KINGDOM.
Product codes and lots
Model #781230. All lots