Z-1015-2013Class II

Pyxis Anesthesia System Model 2000 (PAS 2000)

CareFusion 303, Inc / initiated 2009-11-20 / Terminated / root cause: design

CareFusion Corporation began a recall of Pyxis Anesthesia System Model 2000 (PAS 2000) Product Usage: The PAS 2000 is an anesthesia medication dispensing device that provides medications and inventory... on . FDA classified it as Class II and gives the reason as "The recall was initiated because Carefusion has confirmed that when the PAS 2000 drawers due to inaccessability of drawers caused by system down time or lock-up". As of , the recall status on record is Terminated, and MDM tags the root cause as design.

Recall record

Recall number
Z-1015-2013
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
CareFusion 303, Inc
Product
Pyxis Anesthesia System Model 2000 (PAS 2000) Product Usage: The PAS 2000 is an anesthesia medication dispensing device that provides medications and inventory management for the anesthesiologist at the point of care within the operating room.
Product code
BRY Cabinet, Table And Tray, Anesthesia
Reason for recall
The recall was initiated because Carefusion has confirmed that when the PAS 2000 drawers due to inaccessability of drawers caused by system down time or lock-up.
Root cause tag
design
FDA root cause
Device Design
Action
Carefusion sent a Safety Alert Notification letter dated January 13, 2010 to all affected customers. The letter identified the affected product, problem and actions to be taken. Customers were instructed to ensure back-up procedures were in place in case of system down-time or lock-up, including review of the back-up procedures with staff. The letter included an update to the PAS 2000 User Guide. The letterstates that a CareFusion field representative will be contacting customers to make arrangements to apply warning labels to the PAS 2000 electronics doors. Customers were instructed to complete and return the Response Card included. For questions contact your local field support representative ot the Advocacy department at [email protected].
Quantity in commerce
2637
Distribution
Worldwide Distribution - USA (Nationwide) including CANADA.
Product codes and lots
All Pyxis Anesthesia System Model 2000

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-1015-2013
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-1015-2013Class IITerminatedVoluntary: Firm initiated