Pyxis Anesthesia System Model 2000 (PAS 2000)
CareFusion Corporation began a recall of Pyxis Anesthesia System Model 2000 (PAS 2000) Product Usage: The PAS 2000 is an anesthesia medication dispensing device that provides medications and inventory... on . FDA classified it as Class II and gives the reason as "The recall was initiated because Carefusion has confirmed that when the PAS 2000 drawers due to inaccessability of drawers caused by system down time or lock-up". As of , the recall status on record is Terminated, and MDM tags the root cause as design.
Recall record
- Recall number
- Z-1015-2013
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- CareFusion 303, Inc
- Product
- Pyxis Anesthesia System Model 2000 (PAS 2000) Product Usage: The PAS 2000 is an anesthesia medication dispensing device that provides medications and inventory management for the anesthesiologist at the point of care within the operating room.
- Product code
- BRY Cabinet, Table And Tray, Anesthesia
- Reason for recall
- The recall was initiated because Carefusion has confirmed that when the PAS 2000 drawers due to inaccessability of drawers caused by system down time or lock-up.
- Root cause tag
- design
- FDA root cause
- Device Design
- Action
- Carefusion sent a Safety Alert Notification letter dated January 13, 2010 to all affected customers. The letter identified the affected product, problem and actions to be taken. Customers were instructed to ensure back-up procedures were in place in case of system down-time or lock-up, including review of the back-up procedures with staff. The letter included an update to the PAS 2000 User Guide. The letterstates that a CareFusion field representative will be contacting customers to make arrangements to apply warning labels to the PAS 2000 electronics doors. Customers were instructed to complete and return the Response Card included. For questions contact your local field support representative ot the Advocacy department at [email protected].
- Quantity in commerce
- 2637
- Distribution
- Worldwide Distribution - USA (Nationwide) including CANADA.
- Product codes and lots
- All Pyxis Anesthesia System Model 2000