Medtronic Midas Rex Legend Footed Attachment, AF01. Intended for use with all Legend and Legend EHS motors...
Medtronic Sofamor Danek Usa, Inc - Dallas Distribution began a recall of Medtronic Midas Rex Legend Footed Attachment, AF01. Intended for use with all Legend and Legend EHS motors for general surgery including maxillofacial... on . The reason FDA records is "Component of product attachment may be outside established tolerance with potential for footed portion to break during normal use". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.
Recall record
- Recall number
- Z-0546-2010
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Medtronic Powered Surgical Solutions
- Product
- Medtronic Midas Rex Legend Footed Attachment, AF01. Intended for use with all Legend and Legend EHS motors for general surgery including maxillofacial , craniofacial and sternotomy applications.
- Product code
- HBB Motor, Drill, Pneumatic
- Reason for recall
- Component of product attachment may be outside established tolerance with potential for footed portion to break during normal use.
- Root cause tag
- manufacturing process
- FDA root cause
- Nonconforming Material/Component
- Action
- Medtronic issued an "Urgent: Medical Device Recall" letter dated November 30, 2009. Consignees were notified of the issued and asked to quarantine any affected product and return affected product to the firm for replacement. For further information, contact Medtronic Powered Surgical Solutions Customer Service at 1-800-433-7080 or 1-817-788-6400.
- Quantity in commerce
- 64
- Distribution
- Worldwide Distribution -- Including the United States (DC, TN, NC, KY, AL, FL, NJ, OH, MI, IN, NY, WA, CA, NV, NM, IA, MO, NE and UT), Great Britain, Brazil, Turkey, Singapore and the Netherlands.
- Product codes and lots
- Serial Numbers: A3275-03, A7094-01, B1536-06, B1536-07, B3532-02, B8465-09, C3472-08, C7120-02, C7255-01, D1646-02, D1972-01, D1972-04, D2604-01, D2604-04, D4212-05, D5683-03, D6441-01, D6504-05, P00197407, P00204906, P00360920 - P00360924, P00363252 - P00363256, P00364049 - P00364052, P00365059 - P00365061, P00365786 - P00365792, P00365794, P00365799, P00365800, P00368100 - P00368103, P00369724 - P00369726, P00369728, P00372167 - P00372171, T-1035-07, T2648-02, T3188-06 and T5016-02.