Z-0546-2010

Medtronic Midas Rex Legend Footed Attachment, AF01. Intended for use with all Legend and Legend EHS motors...

Medtronic Powered Surgical Solutions / initiated 2009-11-30 / Terminated / root cause: manufacturing process

Medtronic Sofamor Danek Usa, Inc - Dallas Distribution began a recall of Medtronic Midas Rex Legend Footed Attachment, AF01. Intended for use with all Legend and Legend EHS motors for general surgery including maxillofacial... on . The reason FDA records is "Component of product attachment may be outside established tolerance with potential for footed portion to break during normal use". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-0546-2010
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Medtronic Powered Surgical Solutions
Product
Medtronic Midas Rex Legend Footed Attachment, AF01. Intended for use with all Legend and Legend EHS motors for general surgery including maxillofacial , craniofacial and sternotomy applications.
Product code
HBB Motor, Drill, Pneumatic
Reason for recall
Component of product attachment may be outside established tolerance with potential for footed portion to break during normal use.
Root cause tag
manufacturing process
FDA root cause
Nonconforming Material/Component
Action
Medtronic issued an "Urgent: Medical Device Recall" letter dated November 30, 2009. Consignees were notified of the issued and asked to quarantine any affected product and return affected product to the firm for replacement. For further information, contact Medtronic Powered Surgical Solutions Customer Service at 1-800-433-7080 or 1-817-788-6400.
Quantity in commerce
64
Distribution
Worldwide Distribution -- Including the United States (DC, TN, NC, KY, AL, FL, NJ, OH, MI, IN, NY, WA, CA, NV, NM, IA, MO, NE and UT), Great Britain, Brazil, Turkey, Singapore and the Netherlands.
Product codes and lots
Serial Numbers: A3275-03, A7094-01, B1536-06, B1536-07, B3532-02, B8465-09, C3472-08, C7120-02, C7255-01, D1646-02, D1972-01, D1972-04, D2604-01, D2604-04, D4212-05, D5683-03, D6441-01, D6504-05, P00197407, P00204906, P00360920 - P00360924, P00363252 - P00363256, P00364049 - P00364052, P00365059 - P00365061, P00365786 - P00365792, P00365794, P00365799, P00365800, P00368100 - P00368103, P00369724 - P00369726, P00369728, P00372167 - P00372171, T-1035-07, T2648-02, T3188-06 and T5016-02.