Medical Device
Manufacturing
THE TRADE PAPER OF THE FACTORY FLOOR
Z-1401-2010

V-Probe, Endocare Cryo V Probe, single unit model # CVA2400. The product is used in a minimally invasive...

Varian Medical Systems, Inc. / initiated 2008-12-05 / Terminated

Endocare, a Wholly Owned Subsidiary of HealthTronics, Inc. began a recall of V-Probe, Endocare Cryo V Probe, single unit model # CVA2400. The product is used in a minimally invasive, percutaneous approach to treating prostate cancer in... on . The reason FDA records is "Endocare received a report of a patient death that occurred during a prostate cryoablation procedure. An autopsy attributed the cause of death to an air embolism. The safety notice for the cryoprobes was issued to...". As of , the recall status on record is Terminated.

Recall record

Recall number
Z-1401-2010
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Varian Medical Systems, Inc.
Product
V-Probe, Endocare Cryo V Probe, single unit model # CVA2400. The product is used in a minimally invasive, percutaneous approach to treating prostate cancer in low, moderate and high risk patients. Prostate Cryoablation uses extremely cold temperatures to destroy cancerous tissue.
Product code
GEH Unit, Cryosurgical, Accessories
Reason for recall
Endocare received a report of a patient death that occurred during a prostate cryoablation procedure. An autopsy attributed the cause of death to an air embolism. The safety notice for the cryoprobes was issued to refocus attention on the Indications and Safety Review included with the Physician Training Materials and to reinforce additional safety information.
FDA root cause
Other
Action
The recall was initiated with Endocare sending a notification letter to identified customers on December 5, 2008 and offering an optional replacement unit. The initial correspondence was sent out via an e-mail if the correspondence was returned undeliverable or an e-mail address was not available it was then sent Federal Express to those recipients. The notification letter explained the reason for recall and requested providers refocus their attention on the Indications & Safety Review information contained in the Endocare's Physician Training Materials. Physicians were instructed to contact the firm at 1-949-450-5412 to discuss any questions they may have about the information in the notification letter.
Quantity in commerce
22,741 units total for all products
Distribution
Nationwide.
Product codes and lots
Lot Numbers 08-0539.