Z-0980-2010
Dimension Vista LOCI Reaction Vessel
Siemens Healthcare Diagnostics Inc. / initiated 2009-11-24 / Terminated / root cause: manufacturing process
Siemens Healthcare Diagnostics, Inc. began a recall of Dimension Vista LOCI Reaction Vessel on . The reason FDA records is "Some vessels may leak. If leakage occurs, patient results will be flagged and are not reportable". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.
Recall record
- Recall number
- Z-0980-2010
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Siemens Healthcare Diagnostics Inc.
- Product
- Dimension Vista LOCI Reaction Vessel
- Product code
- JJE Analyzer, Chemistry (Photometric, Discrete), For Clinical Use
- Reason for recall
- Some vessels may leak. If leakage occurs, patient results will be flagged and are not reportable.
- Root cause tag
- manufacturing process
- FDA root cause
- Nonconforming Material/Component
- Action
- The recalling firm issued an Urgent Field Safety Notice dated 11/2009 to all affected customers. Customers have been told to discard any bags of the affected lot. If the affected lot is the only lot the customer has, customers can examine a vessel from the bag. If the vessel mold cavity identity is B8, customers should discard the bag. Thus customers can use the lot if the cavity mold is anything other than B8. In addition, customers have been told to empty the LOCI vessel holder on all systems and to discard vessels in case any affected vessels have been loaded onto the system. For technical questions or concerns, please contact the Siemens Technical Solutions Center at 800-441-9250 for further assistance.
- Quantity in commerce
- 408 units
- Distribution
- The device was shipped to medical facilities in CA, CO, FL, GA, IL, LA, MD, MI, MS, NE, NC, NJ, NY, OH, SC, TN, TX, VA, WA, and WI. The product was also shipped to Belgium, France, Germany, Italy, Netherlands, Portugal, Spain, and Switzerland.
- Product codes and lots
- Catalog number KS855. Lot number NH29-188-09