Z-1583-2010

Stryker Orthopaedics, Duracon 4:1 Ceramic Cutting Block XS; XCelerate cutting block for distal femur in...

Howmedica Osteonics Corp / initiated 2009-12-03 / Terminated / root cause: design

Stryker Howmedica Osteonics Corp. began a recall of Stryker Orthopaedics, Duracon 4:1 Ceramic Cutting Block XS; XCelerate cutting block for distal femur in orthopaedic surgery, Non-Sterile on . The reason FDA records is "The ceramic guide rails, within the XCelerate 4:1 Ceramic Cutting Blocks, may fracture and displace from the block". As of , the recall status on record is Terminated, and MDM tags the root cause as design.

Recall record

Recall number
Z-1583-2010
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Howmedica Osteonics Corp
Product
Stryker Orthopaedics, Duracon 4:1 Ceramic Cutting Block XS; XCelerate cutting block for distal femur in orthopaedic surgery, Non-Sterile.
Product code
LXH Orthopedic Manual Surgical Instrument
Reason for recall
The ceramic guide rails, within the XCelerate 4:1 Ceramic Cutting Blocks, may fracture and displace from the block.
Root cause tag
design
FDA root cause
Device Design
Action
Urgent Product Recall Letters were sent on December 3, 2009 via Federal Express to Stryker Branches/Agencies, Hospital Risk Management, Hospital Chief of Orthopaedics and Surgeons. Foreign accounts were notitied on December 7, 2009. Questions or comments should be addressed to Colleen O'Meara at 201-972-2100.
Quantity in commerce
3056 Duracon Cutting Blocks, total
Distribution
Worldwide Distribution: USA, Sweden, Ireland, Canada, Hong Kong, Germany, Spain, India, Australia, Italy, New Zealand, and United Kingdom.
Product codes and lots
Catalog Number: 8000-7005, all lots