Z-1588-2010
Stryker Orthopaedics, Duracon 4:1 Ceramic Cutting Block XL; XCelerate cutting block for distal femur in...
Howmedica Osteonics Corp / initiated 2009-12-03 / Terminated / root cause: design
Stryker Howmedica Osteonics Corp. began a recall of Stryker Orthopaedics, Duracon 4:1 Ceramic Cutting Block XL; XCelerate cutting block for distal femur in orthopaedic surgery, Non-Sterile on . The reason FDA records is "The ceramic guide rails, within the XCelerate 4:1 Ceramic Cutting Blocks, may fracture and displace from the block". As of , the recall status on record is Terminated, and MDM tags the root cause as design.
Recall record
- Recall number
- Z-1588-2010
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Howmedica Osteonics Corp
- Product
- Stryker Orthopaedics, Duracon 4:1 Ceramic Cutting Block XL; XCelerate cutting block for distal femur in orthopaedic surgery, Non-Sterile.
- Product code
- LXH Orthopedic Manual Surgical Instrument
- Reason for recall
- The ceramic guide rails, within the XCelerate 4:1 Ceramic Cutting Blocks, may fracture and displace from the block.
- Root cause tag
- design
- FDA root cause
- Device Design
- Action
- Urgent Product Recall Letters were sent on December 3, 2009 via Federal Express to Stryker Branches/Agencies, Hospital Risk Management, Hospital Chief of Orthopaedics and Surgeons. Foreign accounts were notitied on December 7, 2009. Questions or comments should be addressed to Colleen O'Meara at 201-972-2100.
- Quantity in commerce
- 3056 Duracon Cutting Blocks, total
- Distribution
- Worldwide Distribution: USA, Sweden, Ireland, Canada, Hong Kong, Germany, Spain, India, Australia, Italy, New Zealand, and United Kingdom.
- Product codes and lots
- Catalog number: 8000-7050, all lots