Z-1035-2010

Pointe Scientific Creatinine Reagent Set, Pointe Scientific, Inc., Canton, MI; Catalog # C7539-150...

Horiba Instruments Incorporated / initiated 2009-12-01 / Terminated / root cause: manufacturing process

Pointe Scientific, Inc. began a recall of Pointe Scientific Creatinine Reagent Set, Pointe Scientific, Inc., Canton, MI; Catalog # C7539-150, C7539-625, C7539-625-S, C7539-1250, HC939-756, HC939-1340... on . The reason FDA records is "An additional 10% picric acid was added to the R2 component during production". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-1035-2010
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Horiba Instruments Incorporated
Product
Pointe Scientific Creatinine Reagent Set, Pointe Scientific, Inc., Canton, MI; Catalog # C7539-150, C7539-625, C7539-625-S, C7539-1250, HC939-756, HC939-1340, 12-C7539-98, 8-C7539-R1-120, 8-C7539-R1-500, 8-C7539-R1-1000, 7-C7539-R1-120, 7-C7539-R1-1000, and 7-C7539-R1-500. For the quantitative determination of creatinine in serum. For in-vitro diagnostic use only.
Product code
CGX Alkaline Picrate, Colorimetry, Creatinine
Reason for recall
An additional 10% picric acid was added to the R2 component during production.
Root cause tag
manufacturing process
FDA root cause
Process change control
Action
Consignees were notified by An Urgent: Medical Device Recall letter, dated 12/1/09, that identified the affected product. Customers were asked to immediately examine their inventory for the affected product and to notify any customers that the product may have been further distributed to by giving them a copy of the recall notice. Customers are to dispose of the affected product immediately. Questions should be directed to the Technical Support Department at (800)757-5313 or (734)487-8300. There is also a Recall Response form attached with this notice.
Quantity in commerce
892 kits
Distribution
Worldwide Distribution -- USA, Bolivia, Greece, Guyana, India, Indonesia, Israel, Kenya, Korea, Malaysia, Mexico, Pakistan, Philippines, Russia, Serbia, Sri Lanka, Tanzania, Trinidad and Ukraine.
Product codes and lots
All lots numbers containing the numbers 923202; exp. 2011-08.