Biomet 2.9 Micromax L-15 suture anchor, flexible long shaft, #2 blue/white maxbraid, PLLA/PGA L-15, sterile...

Zimmer GmbH / initiated 2009-12-11 / Terminated / root cause: labeling and UDI

Biomet, Inc. began a recall of Biomet 2.9 Micromax L-15 suture anchor, flexible long shaft, #2 blue/white maxbraid, PLLA/PGA L-15, sterile, Biomet Sports Medicine, Warsaw, IN; REF 905456... on . The reason FDA records is "The anchor label states it contains a flexible shaft, but this lot actually contians a rigid shaft". As of , the recall status on record is Terminated, and MDM tags the root cause as labeling and UDI.

Recall record

Recall number
Z-0611-2010
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Zimmer GmbH
Product
Biomet 2.9 Micromax L-15 suture anchor, flexible long shaft, #2 blue/white maxbraid, PLLA/PGA L-15, sterile, Biomet Sports Medicine, Warsaw, IN; REF 905456. Intended for use in soft tissue reattachment procedures in the shoulder, wrist, elbow and knee.
Product code
HWC Screw, Fixation, Bone
Reason for recall
The anchor label states it contains a flexible shaft, but this lot actually contians a rigid shaft.
Root cause tag
labeling and UDI
FDA root cause
Labeling mix-ups
Action
Biomet Sports Medicine notified the single consignee by letter dated December 11, 2009. The firm requested that users locate and return the affected product and return a business reply card. For further information, contact Biomet at 1-800-348-9500 or 1-574-372-3983.
Quantity in commerce
6
Distribution
Nationwide Distribution -- Alabama.
Product codes and lots
Lot 695710.