Artisan brand Extend Control Catheter; An Steerable Guide Catheter; Component of Sensei X Robotic Catheter...
Hansen Medical Inc began a recall of Artisan brand Extend Control Catheter; An Steerable Guide Catheter; Component of Sensei X Robotic Catheter System; Model Number: 07546; Distributed and... on . The reason FDA records is "Certain Artisan Extend Catheters may develop a leak in the flush assembly of the catheter which has the potential to cause flush fluid leakage and subsequent decrease in distal flush; inability to aspirate air from...". As of , the recall status on record is Terminated.
Recall record
- Recall number
- Z-1251-2010
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Hansen Medical, Inc.
- Product
- Artisan brand Extend Control Catheter; An Steerable Guide Catheter; Component of Sensei X Robotic Catheter System; Model Number: 07546; Distributed and Manufactured by Hansen Medical, Inc, Mountain View, CA
- Product code
- DXX System, Catheter Control, Steerable
- Reason for recall
- Certain Artisan Extend Catheters may develop a leak in the flush assembly of the catheter which has the potential to cause flush fluid leakage and subsequent decrease in distal flush; inability to aspirate air from flush lines, and/or introduction of air into the catheter.
- FDA root cause
- Other
- Action
- Recall Initiation Date: October 19,2009 (FedEx shipments on October 19, for October 20 delivery) via letters explaining the reason for the recall and requesting the product be returned. Customers with questions are instructed to call the company at 650-404-5940.
- Distribution
- Nationwide to 12 states.
- Product codes and lots
- Serial Numbers: 6394, 6376, 6377, 6378, 6379, 6380, 6381, 6382, 6383, 6384, 6385, 6386, 6387, 6388, 6389, 6390, 6392, 6393, 6395, 6396, 6397, 6398, 6399, 6400, 6401, 6402, 6403, 6406, 6407, 6408, 6409, 6410, 6411, 6412, 6413, 6414, 6415, 6417, 6418, 6419, 6420, 6421, 6422, 6423, 6424, 6425, 6427, 6428, 6429, 6430, 6431, 6432, 6433, 6434, 6435, 6436, 6437, 6438, 6439, 6440, 6441, 6442, 6443, 6444, 6445, 6446, 6447, 6448, 6449, 6450, 6451, 6452, 6453, 6454, 6455, 6456, 6457, 6458, 6459, 6460, 6461, 6462, 6463, 6464, 6465, 6466, 6467, 6468, 6469, 6470, 6471, 6472, 6473, 6474, 6475, 6476, 6477, 6478, 6479, 6480, 6481, 6482, 6483, 6484, 6485, 6486, 6487, 6488, 6489, 6490, 6491, 6492, 6493, 6494, 6495, 6496, 6497, 6498, 6499, 6501, 6502, 6504, 6505, 6506, 6507, 6508, 6509, 6511, 6513, 6514, 6515, 6516, 6517, 6519, 6520, 6522, 6523, 6524, 6525, 6526, 6527, 6528, 6530, 6532, 6533, 6534, 6537, 6542, 6543, 6544, 6549, 6556, 6562, 6564, 6566, 6569, 6570, 6573, 6574, 6575, 6577, 6578, 6580, 6582, 6585, 6586, 6588, 6589, 6591, 6593, 6596, 6598, 6599, 6600, 6601, 6608, 6612, 6615, 6616, 6620, 6621, 6626, 6627, 6633, 6635, 6636, 6638