Z-1329-2010

Lifeshield, latex-free, 150 mL Burette Set, convertible pin, 100 inch with 2 pre-pierced injection sites...

Abbott Medical / initiated 2009-11-24 / Terminated / root cause: design

Hospira Inc. began a recall of Lifeshield, latex-free, 150 mL Burette Set, convertible pin, 100 inch with 2 pre-pierced injection sites, SOLUSET, MICRODRIP, 60 drops/mL, a sterile Rx single... on . The reason FDA records is "Reports of difficulty in adjusting the flow rate, which could result in flow rate variances". As of , the recall status on record is Terminated, and MDM tags the root cause as design.

Recall record

Recall number
Z-1329-2010
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Abbott Medical
Product
Lifeshield, latex-free, 150 mL Burette Set, convertible pin, 100 inch with 2 pre-pierced injection sites, SOLUSET, MICRODRIP, 60 drops/mL, a sterile Rx single use device, Hospira Inc, Lake Forest, IL 60045 USA. made in Costa Rica, list number 11398-20.
Product code
FPA Set, Administration, Intravascular
Reason for recall
Reports of difficulty in adjusting the flow rate, which could result in flow rate variances.
Root cause tag
design
FDA root cause
Component design/selection
Action
Hospira sent the Urgent Device Recall letters dated 11/25/09 to their direct accounts of the affected products and requested these consignees to notify their customers of the recall. The accounts were informed that Hospira ceased manufacturing the I.V. gravity and pump administration sets with the CAIR clamp manufactured by Borla S.P.A. of Italy in March 2008, due to customer reports related to a difference in the zone of control used to adjust the flow compared to the original CAIR clamp design. To avoid the possibility of flow rate variances, Hospira is recalling all remaining sets with the Borla designed CAIR clamp. The accounts were requested to quarantine all inventory of the affected lots, complete the enclosed reply form indicating the amount of product being returned, fax a copy of the completed form to Stericycle at 1-877-377-5114, and include a copy with the returned goods to Stericycle at 2670 Executive Drive, Suite A, Indianapolis, IN 46241.
Quantity in commerce
15,880 sets
Distribution
Worldwide Distribution -- U.S., Puerto Rico, Dominican Republic, Bahamas, Virgin Islands, Argentina, Australia, Bahrain, Belgium, Canada, China, Denmark, Ecuador, EI Salvador, Finland, France, Germany, Greece, Guatemala, Hong Kong, Ireland, Italy, Jordan, Kuwait, Lebanon, Libya, Mexico, Netherlands, New Zealand, Nicaragua, Norway, Oman, Panama, Peru, Philippines, Poland, Portugal, San Marino, Saudi Arabia, Singapore, Slovenia, South Africa, Spain, Sweden, Switzerland, Taiwan, Thailand, Turkey, United Kingdom, Uruguay, and United Arab Emirates.
Product codes and lots
Lot numbers: 561245H, 590405H, 591205H, 591275H, 591485H, 601115H, and 621025H.