Z-0824-2010
Abiomed AB5000 Circulatory Support System, Catalog Number: 0015-0000. Manufactured by Abiomed, Inc. Danvers...
Abiomed, Inc. / initiated 2009-12-16 / Terminated / root cause: manufacturing process
Abiomed, Inc. began a recall of Abiomed AB5000 Circulatory Support System, Catalog Number: 0015-0000. Manufactured by Abiomed, Inc. Danvers, MA. Intended to provide complete short-term... on . The reason FDA records is "Console shut down with no audible alarm". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.
Recall record
- Recall number
- Z-0824-2010
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Abiomed, Inc.
- Product
- Abiomed AB5000 Circulatory Support System, Catalog Number: 0015-0000. Manufactured by Abiomed, Inc. Danvers, MA. Intended to provide complete short-term support of the left and/or right sides of the heart to patients suffering from potentially reversible ventricular dysfunction.
- Product code
- DSQ Ventricular (Assist) Bypass
- Reason for recall
- Console shut down with no audible alarm
- Root cause tag
- manufacturing process
- FDA root cause
- Process control
- Action
- Abiomed initiated the recall by telephone contact and followed with a Recall Letter dated January 27, 2010. Consignees were informed that a Field Service Engineer will be in contact and schedule a time to visit and correct the affected device. For further information, contact Abiomed Quality Assurance at 1-978-656-1543 or Abiomed Field Service at 1-800-554-8666.
- Quantity in commerce
- 496 units
- Distribution
- Worldwide Distribution -- United States, Australia, Brazil, China, Hong Kong, Japan, Mexico and Germany.
- Product codes and lots
- Serial Numbers: AD5001 through AD5496, except AD5006 and AD5018.