Intelligent Hearing Systems Corporation, SmartEP with Vestibular Evoked Myogenic Potentials Version 3.96...
Intelligent Hearing Systems Corp. began a recall of Intelligent Hearing Systems Corporation, SmartEP with Vestibular Evoked Myogenic Potentials Version 3.96. Model M811016 on . The reason FDA records is "This was a marketing correction mandated by the United States Food and Drug Administration (USFDA). USFDA notified Intelligent Hearing Systems that 501(k) premarket notification application must be submitted and cleared...". As of , the recall status on record is Terminated, and MDM tags the root cause as design.
Recall record
- Recall number
- Z-1704-2010
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Intelligent Hearing Systems Corp.
- Product
- Intelligent Hearing Systems Corporation, SmartEP with Vestibular Evoked Myogenic Potentials Version 3.96. Model M811016
- Product code
- GWJ Stimulator, Auditory, Evoked Response
- Reason for recall
- This was a marketing correction mandated by the United States Food and Drug Administration (USFDA). USFDA notified Intelligent Hearing Systems that 501(k) premarket notification application must be submitted and cleared in order to market and sell devices with Vestibular Evoked Myogenic Potential (VEMP) specific functions. The Food and Drug Administration considers that the VEMP testing functions
- Root cause tag
- design
- FDA root cause
- Device Design
- Action
- Intelligent Hearing Systems sent a letter "Information Regarding the Discontinued Use of the SmartEP Evoked Response System for Vestibular Evoked Myogenic Potentials (VEMP) Testing" dated September 30, 2008 to customers and distributors. Each customer will receive an upgrade and VEMP removal kit which will remove the VEMP feature.
- Quantity in commerce
- 50
- Distribution
- Nationwide distribution.
- Product codes and lots
- IHS2195, IHS3048V, IHS3116, IHS3128V, IHS3182, IHS3420, IHS3431, IHS3443, IHS3485, IHS3545, IHS3563, IHS3565, IHS3580, IHS3591, IHS3642, IHS3665, IHS3668, IHS3683, IHS3829, IHS3878, IHS3967, IHS4001, IHS2209, IHS2281, IHS3110, IHS3324, IHS3410, IHS3426, IHS3460, IHS3542, IHS3548, IHS3623, IHS3638, IHS3787, IHS3961, IHS3011, IHS3387, IHS3429, IHS3432, IHS3494, IHS3593, IHS3619, IHS3704, IHS2184, IHS3008, IHS3248, IHS3314, IHS3819, IHS3893, IHS4040.