Stryker Reflex Hybrid Screw Extractor Inner Shaft; Not Sterile; Manufactured by: Stryker Spine SAS, Cestas...
Stryker Spine began a recall of Stryker Reflex Hybrid Screw Extractor Inner Shaft; Not Sterile; Manufactured by: Stryker Spine SAS, Cestas, France; Distributed in the US by: Stryker Spine, 2... on . The reason FDA records is "The threaded tip of the Reflex Hybrid Screw Extractor has been found to fracture the head of the screw, causing problems in removing the screw and possible damage to the screw and/or plate while attempting to remove the...". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.
Recall record
- Recall number
- Z-0763-2010
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Howmedica Osteonics Corp
- Product
- Stryker Reflex Hybrid Screw Extractor Inner Shaft; Not Sterile; Manufactured by: Stryker Spine SAS, Cestas, France; Distributed in the US by: Stryker Spine, 2 Pearl Court, Allendale, NJ. The threaded tip of the Reflex Hybrid Screw Extractor has been found to fracture the head of the screw, causing problems in removing the screw and possible damage to the screw and/or plate while attempting to remove the screw.
- Product code
- KWQ Appliance, Fixation, Spinal Intervertebral Body
- Reason for recall
- The threaded tip of the Reflex Hybrid Screw Extractor has been found to fracture the head of the screw, causing problems in removing the screw and possible damage to the screw and/or plate while attempting to remove the screw.
- Root cause tag
- manufacturing process
- FDA root cause
- Employee error
- Action
- An "Urgent Product Recall" letter was sent to Stryker Branches/Agencies on October 12, 2009 and to surgeons who have used the device on October 26, 2009. The recall letter described the affected products, issue and potential hazards. Consignees were instructed to examine their inventory and hospital locations for the affected products and return to Stryker Spine as indicated in the recall letter. Customers were also instructed to complete the enclosed Customer Response form and fax to 1-201-760-8370. Direct questions about the recall to Stryker Spine by calling 1-201-760-8206.
- Quantity in commerce
- 983
- Distribution
- Worldwide Distribution.
- Product codes and lots
- Ref # 48511905B, all lots.