Acacia Catalog # 385164 Tri-Extension IV Set 15cm Small Bore, Spin Nut w/BD Q-Syte

Medela AG / initiated 2009-12-18 / Terminated / root cause: manufacturing process

Medical Product Specialist Inc dba MPS Acacia began a recall of Acacia Catalog # 385164 Tri-Extension IV Set 15cm Small Bore, Spin Nut w/BD Q-Syte on . The reason FDA records is "Some lots of IV Extension Sets with BD Q-Syte Luer Access Split Septum have a manufacturing deviation which could cause the product to not function properly, possibly resulting in an air embolism". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-1078-2010
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Medela AG
Product
Acacia Catalog # 385164 Tri-Extension IV Set 15cm Small Bore, Spin Nut w/BD Q-Syte
Product code
FPA Set, Administration, Intravascular
Reason for recall
Some lots of IV Extension Sets with BD Q-Syte Luer Access Split Septum have a manufacturing deviation which could cause the product to not function properly, possibly resulting in an air embolism.
Root cause tag
manufacturing process
FDA root cause
Process design
Action
A letter dated December 18, 2009 was sent to Acacia customers with the following instructions. 1. Immediately review your inventory of Acacia IV Extension Sets with BD Q-Syte for REF numbers and lot numbers listed and remove them from all inventory printed on the unit packaging , shelf box and shelf shipper containers. 2. Complete the Distributor Recall response Card (Attachment 1) and return by fax immediately to the number on the Distributor Recall Response Card for product in your facility at the time of receipt of this letter. This form must be completed and f axed to Acacia, Inc even if there is no product in inventory. 3. Acacia, Inc is currently working to ensure that replacement products are readily available. Once you have identified how much product you will be returning, please call your Customer Representative at 1-800-453-4538, opt. 2, ext. 2534, to arrange for product replacement. Questions are to be directed to the Customer Representative at 1-800-453-4538, opt. 2, ext. 2534, between 8:00AM and 5:00PM MST, Monday-Friday.
Quantity in commerce
6,936
Distribution
US Only
Product codes and lots
Lot A1951