Optovue RTVue Optical Coherence Tomography (OCT), Model-RT100 with software versions 3.0.x.x and higher, with...
Optovue Inc. began a recall of Optovue RTVue Optical Coherence Tomography (OCT), Model-RT100 with software versions 3.0.x.x and higher, with Normative Database, manufactured by Optovue... on . The reason FDA records is "Marketed without proper 510(k)-- No 510(k) clearance for Optovue RTVue Model-RT100 with Software Versions 3.0.x.x and higher which includes the Normative Database". As of , the recall status on record is Terminated, and MDM tags the root cause as software.
Recall record
- Recall number
- Z-1131-2010
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Optovue Inc.
- Product
- Optovue RTVue Optical Coherence Tomography (OCT), Model-RT100 with software versions 3.0.x.x and higher, with Normative Database, manufactured by Optovue, Fremont, CA The RTVue is an optical coherence tomography system indicated for the in vivo imaging and measurement of the retina, retinal nerve fiber layer, optic disk, and for cornea and anterior eye scans as an aid in the diagnosis and management of retinal disease.
- Product code
- OBO Tomography, Optical Coherence
- Reason for recall
- Marketed without proper 510(k)-- No 510(k) clearance for Optovue RTVue Model-RT100 with Software Versions 3.0.x.x and higher which includes the Normative Database.
- Root cause tag
- software
- FDA root cause
- Other
- Action
- The firm is preparing a software "Roll Back" version of software to disable the NDB and other features, ensuring that this version will work properly. A customer notification letter is provided as well as a correction verification form, installation instructions and user manual. If there are any questions, please call 1-866-344-8948.
- Quantity in commerce
- 1800 units
- Distribution
- Nationwide and Worldwide
- Product codes and lots
- Serial Numbers from 10005 to 11805 (serial number are sequential)
FIRST ARTICLE / WEEKLY