Encore Medical Foundation Knee Revision Baseplate, Part Number 321-01-106, Size 6, Left
Encore Medical, Lp began a recall of Encore Medical Foundation Knee Revision Baseplate, Part Number 321-01-106, Size 6, Left on . The reason FDA records is "Partial lot of product was not manufactured to specification. Central locking screw was missing threads. There is a potential for dissociation because the screw is not able to be used as a secondary fixation method". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.
Recall record
- Recall number
- Z-0826-2010
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Encore Medical, LP
- Product
- Encore Medical Foundation Knee Revision Baseplate, Part Number 321-01-106, Size 6, Left
- Product code
- JWH Prosthesis, Knee, Patellofemorotibial, Semi-Constrained, Cemented, Polymer/Metal/Polymer
- Reason for recall
- Partial lot of product was not manufactured to specification. Central locking screw was missing threads. There is a potential for dissociation because the screw is not able to be used as a secondary fixation method.
- Root cause tag
- manufacturing process
- FDA root cause
- Process control
- Action
- Firm notified consignees by letter on 12/11/2009. The letter informs physicians that patients with implanted devices should counseled about the issue and should report any changes to their treating physician.
- Quantity in commerce
- 17
- Distribution
- TX, FL, GA, CA, AZ, Japan
- Product codes and lots
- Lot 53944312