Medical Device
Manufacturing
THE TRADE PAPER OF THE FACTORY FLOOR
Z-0826-2010

Encore Medical Foundation Knee Revision Baseplate, Part Number 321-01-106, Size 6, Left

Encore Medical, LP / initiated 2009-12-11 / Terminated / root cause: manufacturing process

Encore Medical, Lp began a recall of Encore Medical Foundation Knee Revision Baseplate, Part Number 321-01-106, Size 6, Left on . The reason FDA records is "Partial lot of product was not manufactured to specification. Central locking screw was missing threads. There is a potential for dissociation because the screw is not able to be used as a secondary fixation method". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-0826-2010
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Encore Medical, LP
Product
Encore Medical Foundation Knee Revision Baseplate, Part Number 321-01-106, Size 6, Left
Product code
JWH Prosthesis, Knee, Patellofemorotibial, Semi-Constrained, Cemented, Polymer/Metal/Polymer
Reason for recall
Partial lot of product was not manufactured to specification. Central locking screw was missing threads. There is a potential for dissociation because the screw is not able to be used as a secondary fixation method.
Root cause tag
manufacturing process
FDA root cause
Process control
Action
Firm notified consignees by letter on 12/11/2009. The letter informs physicians that patients with implanted devices should counseled about the issue and should report any changes to their treating physician.
Quantity in commerce
17
Distribution
TX, FL, GA, CA, AZ, Japan
Product codes and lots
Lot 53944312