Z-0796-2010

Depth Gauge; Catalog Number: 703707; Stryker Trauma AG, CH 2545 Selzach; Distributed in USA by Howmedica...

Howmedica Osteonics Corp / initiated 2009-12-22 / Terminated / root cause: manufacturing process

Stryker Howmedica Osteonics Corp. began a recall of Depth Gauge; Catalog Number: 703707; Stryker Trauma AG, CH 2545 Selzach; Distributed in USA by Howmedica Osteonics Corp., 325 Corporate Drive, Mahwah, NJ... on . The reason FDA records is "The Depth Measuring Gauges have 5 scale marking instead of the required 4 scale markings". As of , the recall status on record is Terminated, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-0796-2010
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Howmedica Osteonics Corp
Product
Depth Gauge; Catalog Number: 703707; Stryker Trauma AG, CH 2545 Selzach; Distributed in USA by Howmedica Osteonics Corp., 325 Corporate Drive, Mahwah, NJ 07430. This product is sold separately or as part of Kit # 990256. The depth gauge is an instrument in the VariAx Elbow System. Depth gauges are tools used intra-operatively, to aid in correct screw selection. They can be used with a plate, or directly on bone.
Product code
HTJ Gauge, Depth
Reason for recall
The Depth Measuring Gauges have 5 scale marking instead of the required 4 scale markings.
Root cause tag
manufacturing process
FDA root cause
Process control
Action
Urgent Product Recall letters, dated December 22, 2009, were sent to all Stryker Branches/Agencies. Letters were also sent to hospitals and surgeons that use the VariaAx Elbow System were also notified. All letters were sent by Federal Express. The letters identified the affected product, stated the issue, and the hazards involved. It also asked customers to examine their inventory and hospital locations to identify affected products. All affected products are to be retrieved and returned to their branch or agency warehouse for reconciliation. All products should be reconciled on the attached Product Accountability Form and a copy should be faxed to 201-831-6069 within 5 days of receipt of the notice. Follow the instructions for shipping the affected product back to the firm. Questions should be directed to 201-972-2100.
Quantity in commerce
49
Distribution
Nationwide Distribution
Product codes and lots
Lot code: U09984  VariaAx Elbow System Instrument Set - kitted, material transaction lots # U23151, U24021, U25906 and U29985 all contain one piece of the affected recalled catalog number.