Exel Safety Huber Infusion Sets (20G and 22G sizes with Catalog Numbers 37854S, 37855S. 37858S)
Exelint International Co. began a recall of Exel Safety Huber Infusion Sets (20G and 22G sizes with Catalog Numbers 37854S, 37855S. 37858S) on . The reason FDA records is "Coring of port's septum, which may lead to: Decreased effectiveness of port, infection, necrosis, swelling, and embolization, which may result in Pulmonary Embolism, Limb ischemia, stroke, local extravasation of...". As of , the recall status on record is Terminated, and MDM tags the root cause as design.
Recall record
- Recall number
- Z-0639-2010
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Exelint International Co.
- Product
- Exel Safety Huber Infusion Sets (20G and 22G sizes with Catalog Numbers 37854S, 37855S. 37858S).
- Product code
- FPA Set, Administration, Intravascular
- Reason for recall
- Coring of port's septum, which may lead to: Decreased effectiveness of port, infection, necrosis, swelling, and embolization, which may result in Pulmonary Embolism, Limb ischemia, stroke, local extravasation of medication such as chemotherapy or other caustic material, may lead to local tissue necrosis, granulomas, failure to deliver therapy, need for additional surgery to replace implanted port
- Root cause tag
- design
- FDA root cause
- Device Design
- Action
- Distributors have been advised to cease distribution of any recalled needles and infusion sets; to quarantine any remaining; to document amount on Customer Response Form; to return form to Exel (firm); to arrange for return of product to Exel for credit. Distributors also are advised to share recall notice with all affected customers. Contact phone: 727-827-1922x0 (M-F 9a-5p EST). Contact email: [email protected]
- Quantity in commerce
- 3,360
- Distribution
- US (VA TX NV KY OH NH MS NY OH NM DE WA AZ IL PA OR CO FL GA LA UT MI WI NV CA NJ MO IA) and Canada.
- Product codes and lots
- Catalog 37854S Lot 09G31; Catalog 37855S Lot 09G31, Catalog 37858S Lot 09G31.
FIRST ARTICLE / WEEKLY