Z-1040-2010

Teleflex Medical, Hudson RCI, Concha Therm Neptune, Heated Humidifier. Product numbers: 425-00, 425-10 and...

Teleflex Medical / initiated 2010-01-12 / Terminated / root cause: design

Telefelx Medical began a recall of Teleflex Medical, Hudson RCI, Concha Therm Neptune, Heated Humidifier. Product numbers: 425-00, 425-10 and 525-30 on . The reason FDA records is "When turned on, the Neptune heater may cause an artifact on the patient's heart monitor". As of , the recall status on record is Terminated, and MDM tags the root cause as design.

Recall record

Recall number
Z-1040-2010
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Teleflex Medical
Product
Teleflex Medical, Hudson RCI, Concha Therm Neptune, Heated Humidifier. Product numbers: 425-00, 425-10 and 525-30.
Product code
BTT Humidifier, Respiratory Gas, (Direct Patient Interface)
Reason for recall
When turned on, the Neptune heater may cause an artifact on the patient's heart monitor.
Root cause tag
design
FDA root cause
Device Design
Action
Teleflex consignees were notified via Urgent Medical Device Recall letter via next day delivery on/about 01/12/2010. They were asked to return all of the affected product. Each consignee that is a distributor was instructed to forward the letter to their consignees to retrieve relevant product from those locations. A second mailing to non-responders on the consignee list will be made approximately four to six weeks after the first mailing. A third mailing to non-responding customers on the consignee list will be made approximately four to six weeks after the second mailing or as an alternative, they may be contacted by phone, e-mail or fax. A record of these contacts will be maintained. Affected products received by Teleflex Medical will be reworked with software upgrade and placed back into distribution. Consignees will either receive a software upgrade on site by a third party or replacement product with the software upgrade. Records will be maintained accordingly.
Quantity in commerce
15,134 units
Distribution
U.S. Canada, India, Japan, Puerto Rico, Germany, France, Italy and Ireland.
Product codes and lots
Product numbers: 425-00, 425-10 and 525-30, All lot and serial numbers.