Z-1040-2010
Teleflex Medical, Hudson RCI, Concha Therm Neptune, Heated Humidifier. Product numbers: 425-00, 425-10 and...
Teleflex Medical / initiated 2010-01-12 / Terminated / root cause: design
Telefelx Medical began a recall of Teleflex Medical, Hudson RCI, Concha Therm Neptune, Heated Humidifier. Product numbers: 425-00, 425-10 and 525-30 on . The reason FDA records is "When turned on, the Neptune heater may cause an artifact on the patient's heart monitor". As of , the recall status on record is Terminated, and MDM tags the root cause as design.
Recall record
- Recall number
- Z-1040-2010
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Teleflex Medical
- Product
- Teleflex Medical, Hudson RCI, Concha Therm Neptune, Heated Humidifier. Product numbers: 425-00, 425-10 and 525-30.
- Product code
- BTT Humidifier, Respiratory Gas, (Direct Patient Interface)
- Reason for recall
- When turned on, the Neptune heater may cause an artifact on the patient's heart monitor.
- Root cause tag
- design
- FDA root cause
- Device Design
- Action
- Teleflex consignees were notified via Urgent Medical Device Recall letter via next day delivery on/about 01/12/2010. They were asked to return all of the affected product. Each consignee that is a distributor was instructed to forward the letter to their consignees to retrieve relevant product from those locations. A second mailing to non-responders on the consignee list will be made approximately four to six weeks after the first mailing. A third mailing to non-responding customers on the consignee list will be made approximately four to six weeks after the second mailing or as an alternative, they may be contacted by phone, e-mail or fax. A record of these contacts will be maintained. Affected products received by Teleflex Medical will be reworked with software upgrade and placed back into distribution. Consignees will either receive a software upgrade on site by a third party or replacement product with the software upgrade. Records will be maintained accordingly.
- Quantity in commerce
- 15,134 units
- Distribution
- U.S. Canada, India, Japan, Puerto Rico, Germany, France, Italy and Ireland.
- Product codes and lots
- Product numbers: 425-00, 425-10 and 525-30, All lot and serial numbers.