Z-0825-2010

Synthes In Situ Bender/Cutter Kit-Wide-Sterile. Catalog Number 530.521S

Synthes GmbH / initiated 2010-01-06 / Terminated

Synthes USA (HQ), Inc. began a recall of Synthes In Situ Bender/Cutter Kit-Wide-Sterile. Catalog Number 530.521S on . The reason FDA records is "There is the potential for the bender/cutter attachment to continue heating after release of the power button. There is potential for patient and/or staff injury if continuous heating is unrecognized". As of , the recall status on record is Terminated.

Recall record

Recall number
Z-0825-2010
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Synthes GmbH
Product
Synthes In Situ Bender/Cutter Kit-Wide-Sterile. Catalog Number 530.521S.
Product code
HRS Plate, Fixation, Bone
Reason for recall
There is the potential for the bender/cutter attachment to continue heating after release of the power button. There is potential for patient and/or staff injury if continuous heating is unrecognized.
FDA root cause
Other
Action
An "Urgent: Medical Device Recall" letter dated January 8, 2010 was issued to customers via USPS Certified mail. The letter described the issue and affected product. Customers were notified to cease use of the recalled product, examine inventory and follow instructions for return to the firm as indicated in the recall letter. For questions, please call 1-800-620-7025 x 5452 or 1-610-719-5452 or contact your Synthes CMF Sales Consultant.
Quantity in commerce
187 units
Distribution
Nationwide Distribution.
Product codes and lots
All lots.