Medical Device
Manufacturing
THE TRADE PAPER OF THE FACTORY FLOOR
Z-1081-2010

Suturable DuraGen Dural Regeneration Matrix; Non-Pyrogenic; Sterility guaranteed unless package is damaged or...

Integra LifeSciences Corporation / initiated 2010-01-13 / Terminated

Integra LifeSciences Corp began a recall of Suturable DuraGen Dural Regeneration Matrix; Non-Pyrogenic; Sterility guaranteed unless package is damaged or opened. Rx only. Manufacturer: Integra... on . The reason FDA records is "One lot of Suturable DuraGen Dural Regeneration Matrix (# 1091575) did not pass stability testing at certain time points for Suture Retention Strength and Shrink Temperature specifications". As of , the recall status on record is Terminated.

Recall record

Recall number
Z-1081-2010
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Integra LifeSciences Corporation
Product
Suturable DuraGen Dural Regeneration Matrix; Non-Pyrogenic; Sterility guaranteed unless package is damaged or opened. Rx only. Manufacturer: Integra LifeSciences Corporation, Plainsboro, NJ 08536 Suturable DuraGen is an absorbable implant for repair of dural defects. It is indicated as a dural substitute for the repair of dura matter.
Product code
GXQ Dura Substitute
Reason for recall
One lot of Suturable DuraGen Dural Regeneration Matrix (# 1091575) did not pass stability testing at certain time points for Suture Retention Strength and Shrink Temperature specifications.
FDA root cause
Other
Action
An "URGENT PRODUCT RECALL NOTIFICATION" dated January 13, 2010, was sent via Federal Express to customers. The notification described the product, problem, hazard involved and action to be taken by customers. In the event that the customers have in their possession any products from the specific lot number 1091576, the customers are being ask to please immediately quarantine all stock on hand and contact their local Integra Sales representative for further assistance and for arrangements of product return. The customers are being ask to return the enclosed "Recall Acknowledge and Return Form" indicating whether or not they possess any of the affected product identified. Integra will replace any product affected by this Recall that is returned. Please feel free to contact me at 609-936-5407 should you have any additional questions.
Quantity in commerce
94
Distribution
Nationwide distribution:Ca, NY, TX, CO, VA, SC, IA, MO, IL, LA, OR, OH, WI, FL, DC, DE, MD, MN.
Product codes and lots
Catalog Number DURS3391; Lot number 1091575