Z-1096-2010

Smart Perfusion Pack, REF: 088504500, Sterile EO, Sorin Group USA, Inc., 14401 W. 65th Way, Arvada, CO 80004...

LivaNova USA, Inc. / initiated 2009-12-30 / Terminated / root cause: packaging

Sorin Group USA, Inc. began a recall of Smart Perfusion Pack, REF: 088504500, Sterile EO, Sorin Group USA, Inc., 14401 W. 65th Way, Arvada, CO 80004 Indicated for use in surgical procedures requiring... on . The reason FDA records is "Tubing set packaging seal between lid and tray may not be adequately sealed, compromosing sterility". As of , the recall status on record is Terminated, and MDM tags the root cause as packaging.

Recall record

Recall number
Z-1096-2010
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
LivaNova USA, Inc.
Product
Smart Perfusion Pack, REF: 088504500, Sterile EO, Sorin Group USA, Inc., 14401 W. 65th Way, Arvada, CO 80004 Indicated for use in surgical procedures requiring extracorporeal support for periods of up to six hours.
Product code
DWF Catheter, Cannula And Tubing, Vascular, Cardiopulmonary Bypass
Reason for recall
Tubing set packaging seal between lid and tray may not be adequately sealed, compromosing sterility.
Root cause tag
packaging
FDA root cause
Packaging process control
Action
Customers were notified by letter on December 30, 2009 and asked to examine all inventory for seal voids and arrange for return if any were found defective. Examination instructions were included. The two distributors were given the same instructions and were told to share all the information with any of their customers.
Quantity in commerce
17 units
Distribution
AZ, CA, FL, MO, NJ, SC, SD, TN, TX, UT, VA and WI.
Product codes and lots
Lot Number: 0932800048.